From the paper's conclusion
  • Device-related issues are primarily intra-operative and lead to procedural disruption with minimal patient harm, whereas non-device-related complications are more common but generally mild. These findings support careful pre-operative preparation, patient selection and counselling.

Abstract

BACKGROUND: Benign prostatic hyperplasia (BPH) is a common condition among men. Minimally invasive surgical therapies (MISTs), including Rezūm (water vapour thermal therapy), are increasingly used due to reduced morbidity and effective symptom relief. However, device-related issues are not routinely captured in clinical studies. This study aimed to analyse adverse events related to Rezūm therapy reported in the MAUDE database. METHODS: A retrospective review of the MAUDE database identified adverse events related to Rezūm therapy between 2017 and 2026. Data were extracted on event characteristics, timing, classification (device vs non-device), clinical impact and manufacturer assessment. After exclusion of duplicates and incomplete reports, 610 adverse events were analysed. RESULTS: Device-related events accounted for 271 cases, most commonly generator or system malfunction (52.4%), occurring predominantly intra-operatively (92.6%) and leading to procedure abandonment in 62.7%. Manufacturer-confirmed faults were reported in 21.8%. Non-device-related events were more frequent (339 cases) and mainly post-operative (90.2%). The most common complications were irritative lower urinary tract symptoms (19.5%), urinary tract infection (10.6%), and urinary retention (acute 5.6%, persistent 2.7%). Bleeding was usually mild, and sexual dysfunction was infrequent. Most procedures were completed (84%), with abandonment in 14.5%. Major complications were rare. CONCLUSIONS: Device-related issues are primarily intra-operative and lead to procedural disruption with minimal patient harm, whereas non-device-related complications are more common but generally mild. These findings support careful pre-operative preparation, patient selection and counselling.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers