SkinFibroMeter in lower extremity lymphedema assessment: Reliability, correlation with ultrasound, and the determination of minimal detectable change.
Elif Duygu-Yildiz, Samet Genez, Serkan Sevim, Beyza Özcan, Gozde Tekin, Nuriye Özengin
Faculty of Health Sciences, Department of Physiotherapy and Rehabilitation, Bolu Abant Izzet Baysal University, Bolu, Turkey. · Faculty of Medicine, Department of Radiology, Bolu Abant Izzet Baysal University, Bolu, Turkey.
Key points
From the paper's conclusion- The secondary aim was to examine its correlation with skin/subcutaneous tissue thickness and establish minimal detectable change (MDC95) values.MethodsThirty lower extremities with physician-diagnosed lymphedema were included in this cross-sectional study.
Abstract
ObjectivesThis study aimed to determine the inter- and intra-rater reliabilities of the SkinFibroMeter in individuals with lower extremity lymphedema. The secondary aim was to examine its correlation with skin/subcutaneous tissue thickness and establish minimal detectable change (MDC95) values.MethodsThirty lower extremities with physician-diagnosed lymphedema were included in this cross-sectional study. Tissue stiffness was assessed at standardized anatomical sites using the SkinFibroMeter® by four raters with varying clinical experience. Skin and subcutaneous tissue thickness were measured at the same points via ultrasound (US). Inter- and intra-rater reliabilities were evaluated using intraclass correlation coefficients (ICC), and the standard error of measurement and MDC95 were calculated.ResultsParticipants had a mean age of 68 ± 11.9 years, a mean BMI of 34.0 ± 6.0 kg/m2, and a mean limb volume of 10,020 ± 3,148 mL. Intra-rater reliability ranged from moderate to excellent (ICC = 0.600-0.976), while inter-rater reliability varied by site (ICC = 0.647-0.941), with the highest agreement at 20 cm above the patella and the lowest at the foot dorsum. Bland-Altman analysis confirmed acceptable absolute agreement across rater pairs with negligible systematic bias. Low-to-moderate positive correlations were observed between the indentation values of the SkinFibroMeter and skin thickness (r = 0.377-0.556, p < .05), whereas negative correlations were found with subcutaneous thickness at several points (r = -0.508 to -0.421, p < .05). No statistically significant correlation was found between localized stiffness and total limb volume (r = -0.023, p = .907). MDC95 values ranged from 0.013 to 0.158.ConclusionThe SkinFibroMeter demonstrates acceptable clinical reliability, with values representing a complex interplay of dermal thickness, interstitial hydration, and tissue compliance rather than global limb size or pure fibrosis. This is the first study to establish MDC95 thresholds, providing clinicians with reference baselines for treatment-related changes. Future studies should incorporate larger, stage-homogeneous cohorts using 3D scanning and shear-wave elastography for comprehensive tissue characterization.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Lower limb lymphedema: reliability of four non-imaging measurement modalities.Disability and rehabilitation · 2026
- Novel Ultrasound Technique for Assessing Fibrosis in Subcutaneous Fat in Lymphedematous Limbs.Lymphatic research and biology · 2026
- Quantitative ultrasound skin thickness measurement for assessment of lower-limb lymphedema.Journal of plastic, reconstructive & aesthetic surgery : JPRAS · 2026
- Do not mistake symptomatic improvement for disease modification: therapeutic endpoints in secondary lower-extremity lymphedema require inflammation-fibrosis anchoring.Frontiers in immunology · 2026
- Layer-Specific Assessment of Dermal Thickness and Stiffness Using Shear Wave Elastography in Gynecologic Cancer-Related Lower-Limb Lymphedema: An Exploratory Cross-Sectional Study.International journal of general medicine · 2026
- Multimodal Assessment of Breast Cancer-Related Lymphedema Using Bioimpedance Analysis, Ultrasound Tissue Thickness, and Shear Wave Elastography.Diagnostics (Basel, Switzerland) · 2026