科研速览 · Science Skim继续刷下去 · Keep skimming →
◆ Ultrasonic imaging2026-08-10

Ultrasound B-Scan to Thru-Scan Reconstructions of Scatterers via Power Law Constraints.

Joaquin Sanchez, Stefano E Romero, Kevin J Parker

原始摘要(英文原文)· Original abstract
This work presents the first application of the Thru-scan framework to a range of ultrasound images across frequency bands, from 3 MHz human abdominal scans to high frequency preclinical scans. The Thru-scan is a constrained reconstruction of the subresolved scatterers based on a power law distribution and enforced by appropriate selection of power law filters. The resulting reconstructions mitigate the traditional speckle "noise" pattern, with substantially increased resolution at high spatial frequencies. Improvements in contrast and apparent resolution are quantified using the generalized contrast-to-noise ratio (gCNR): Thru-scan increases gCNR by approximately 0.07 to 0.10 units, with p-values of less than 0.001. The Thru-scan processing steps also modify the amplitude distribution of the traditional B-scan, and across all frequency bands and anatomical targets the results are well modeled as a Burr distribution with power law exponents near 2.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Ultrasound B-Scan to Thru-Scan Reconstructions of Scatterers via Power Law Constraints. — 科研速览 Science Skim