Preparation and Evaluation of In-House Multi-Analyte Quality Control Materials for ANA, Anti-dsDNA, and Autoantibody Profile Assays.
Lin Xu, Hongxiang Xie
Clinical LaboratoryAffiliated Hospital of Shaoxing University.
Key points
From the paper's conclusion- The in-house QC materials showed good stability and reproducibility, with simple preparation and potential suitability for routine use in primary healthcare laboratories.
Abstract
BACKGROUND: This study explored the preparation and evaluation of in-house quality control (QC) materials for antinuclear antibody (ANA) and anti-double-stranded DNA antibody (anti-dsDNA) testing by indirect immunofluorescence test (IIFT), and ANA profile testing by line immunoassay (LIA), with the aim of providing a reference for other in-house QC materials. METHODS: Patient sera were pooled to prepare a QC serum positive for all targets, while sera from healthy individuals were pooled to prepare a negative control. QC materials were heat-inactivated, centrifuged, aliquoted, and stored at -20 °C. Stability was assessed at 37 °C and during long-term storage at -20 °C. Inter-vial variation was evaluated using 20 vials, and intra-vial variation was assessed by 20 replicates of one vial. RESULTS: During the 8-day 37 °C stability test, ANA and anti-dsDNA titers remained within ± 1 serial dilution of the baseline value, and all ANA profile antibodies remained within the predefined ± 20% acceptance criterion. During long-term storage, ANA and anti-dsDNA titers also remained within ± 1 serial dilution, and the CVs of all ANA profile antibodies were below 20%. Both inter-vial and intra-vial variation met predefined criteria, and all negative controls remained negative. CONCLUSIONS: The in-house QC materials showed good stability and reproducibility, with simple preparation and potential suitability for routine use in primary healthcare laboratories.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Analytical Comparison of an In-House FARR Fluorescence Immunoassay with Commercial Analytical Assays for the Detection of Anti-Double-Stranded DNA Antibodies.Diagnostics (Basel, Switzerland) · 2026
- Long-term stability and reproducibility of frozen plasma control samples for hepatitis B virus, hepatitis C virus, and human immunodeficiency virus by quantitative PCR: Implications for retrospective testing in hemovigilance.Transfusion · 2026
- A-183 Standardizing Specimen Stability Studies for Venous Whole Blood, Serum, and Plasma: A Practical Framework with Analysis Methods, Acceptance Criteria, and Worked ExamplesClinical Chemistry · 2026
- Refrigerated storage stability of 11 humoral parameters in two different venous blood collection tubes over a 7-day period.Clinical chemistry and laboratory medicine · 2026
- Extended in-use stability of Krebs von den Lungen-6 (KL-6) reagents, quality controls, and calibrators under refrigerated conditionsInternational Medical Research Journal · 2026
- Evaluation and application of automated quality control in clinical coagulation testing.Research and practice in thrombosis and haemostasis · 2026