Cloud-Like Thalamic Edema After Redo Magnetic Resonance-Guided Focused Ultrasound Ventral Intermediate Nucleus Thalamotomy: A Case Report.
Kun-Hong Li, Pei-Han Wu, Wei-Chieh Chang
MR-guided Focused Ultrasound Center, Chang Bing Show Chwan Memorial Hospital, Changhua, Taiwan.
Key points
From the paper's conclusion- The most plausible pathogenesis is mechanically mediated injury from cavitation-related acoustic activity. We hypothesize that substantial overlap with the prior ablation zone was the primary predisposing factor, potentially augmented by calcification-related scattering, No-Pass Region (NPR)-driven aperture distortion, occult skull marrow changes, and standing-wave formation. We discuss biophysical mechanisms and propose physics-guided safeguards for redo procedures.
Abstract
INTRODUCTION: Cavitation-associated adverse events in magnetic resonance-guided focused ultrasound (MRgFUS) thalamotomy are uncommon but potentially serious. We report a case of acute "electric-shock" sensation during an alignment/verification sonication followed by thalamic T2 hyperintensity/edema after a redo ventral intermediate nucleus (VIM) thalamotomy. We discuss plausible biophysical mechanisms and risk-reduction strategies. CASE PRESENTATION: A 53-year-old woman with essential tremor underwent repeat MRgFUS 5 months after her initial treatment. During the 4th alignment/verification sonication (planned 200 W, 20 s; measured power approximately 192 W), cavitation monitoring triggered a halt at 19.2 s after 2,037 J was delivered. Despite a maximum target temperature <50°C, the patient immediately reported an "electric-shock" sensation. Post-procedure, she developed right-sided sensory symptoms, mild dysarthria, and hypogeusia. Early magnetic resonance imaging (MRI) showed irregular thalamic hyperintensity consistent with edema, but without confirmed off-target thermal rise on intraoperative thermometry. CONCLUSION: The most plausible pathogenesis is mechanically mediated injury from cavitation-related acoustic activity. We hypothesize that substantial overlap with the prior ablation zone was the primary predisposing factor, potentially augmented by calcification-related scattering, no-pass region-driven aperture distortion, occult skull marrow changes, and standing-wave formation. We discuss biophysical mechanisms and propose physics-guided safeguards for redo procedures.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Staged bilateral MR-guided focused ultrasound thalamotomy in a patient with cerebral cavernous malformations.BMJ case reports · 2026
- Safety of Peri-procedural antithrombotic agents in magnetic resonance-guided focused ultrasound thalamotomy.Clinical neurology and neurosurgery · 2026
- Safety of Staged Bilateral MRgFUS Thalamotomy: Associations Between Lesion Characteristics, Stereotactic Coordinates, and Adverse Events.Neurosurgery · 2026
- Ultrasound Modulation of Brain Function.The Neuroscientist : a review journal bringing neurobiology, neurology and psychiatry · 2026
- Is It Safe to Have Deep Brain Stimulation after MRI-Guided Focused Ultrasound Ablation? A Report of Three Cases and Review of Literature.Stereotactic and functional neurosurgery · 2026
- MR-Guided Focused Ultrasound in Neurology: Clinical Evidence and the Emerging Role of Artificial Intelligence—A Narrative ReviewAI · 2026