Needle-based Confocal Laser Endomicroscopy (nCLE) for Tool-in-Lesion Confirmation in Cone-Beam CT-guided Navigation Bronchoscopy: a Prospective Concordance Study.
Saskia van Heumen, Onno M Mets, Daniel M de Bruin, Peter I Bonta, Johannes M A Daniels, Marjolein A Heuvelmans, Jouke T Annema
Key points
From the paper's conclusion- nCLE and CBCT offer complementary tool position information, with nCLE delivering microscopic-level insights and CBCT providing three-dimensional anatomical context. In environments where CBCT is not available, nCLE may serve as an alternative.
Abstract
BACKGROUND AND OBJECTIVE: Positioning sampling instruments within pulmonary nodules remains a challenge during navigation bronchoscopy. Cone-beam computed tomography (CBCT) and needle-based confocal laser endomicroscopy (nCLE) enable tool positioning feedback. However, access to CBCT systems is limited and associated with radiation exposure. To date, no direct comparison between these two modalities has been reported. This study compares CBCT and nCLE for sampling tool position in navigation bronchoscopy for peripheral pulmonary nodules (PPNs). METHODS: Twenty-five patients with PPNs scheduled for CBCT-guided navigation bronchoscopy were prospectively enrolled. nCLE imaging, CBCT image acquisition and needle aspiration were performed twice for each nodule, followed by (cryo)biopsies. nCLE videos were evaluated using validated criteria and categorized as tool-in-lesion (TIL) or tool-out-lesion (TOL). CBCT scans were classified as TIL, tool-in-contact (TIC), or TOL. RESULTS: nCLE imaging was performed in 24 patients (nodule was not reached in one), with a median nodule diameter of 19 mm, yielding good-quality images in 22/24 patients (92%). In 77% of these cases, there was an agreement between nCLE and CBCT on the needle position. There were no cases in which CBCT confirmed TIL while nCLE did not. From the five discrepant cases, four were TIC on CBCT, while nCLE was TOL. In one case, there was a complete discrepancy between the techniques with nCLE as TIL and CBCT TOL. CONCLUSION: nCLE and CBCT offer complementary tool position information, with nCLE delivering microscopic-level insights and CBCT providing three-dimensional anatomical context. In environments where CBCT is not available, nCLE may serve as an alternative.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Efficacy and safety of the RC120 navigation system for the computed tomography-guided localization of pulmonary ground-glass nodules: a prospective study with historical controls.Quantitative imaging in medicine and surgery · 2026
- Navigation Systems Significantly Improve the Efficiency and Safety of CT-Guided InterventionsLife · 2026
- Cone-beam computed tomography augmented-fluoroscopy-guided center-strike targeting during peripheral bronchoscopy: robotic vs. manual catheters.Journal of thoracic disease · 2026
- Tool in lesion verification of shape-sensing robotic-assisted bronchoscopy with cone beam CT in sampling peripheral pulmonary nodulesThorax · 2025
- Electromagnetic bronchoscopy versus computed tomography for pulmonary nodule localization.Interdisciplinary cardiovascular and thoracic surgery · 2026
- Comparison of diagnostic yield between electromagnetic navigation bronchoscopy-guided forceps biopsy and target-fixing technique-assisted RP-EBUS-guided transbronchial cryobiopsy with a guide sheath for peripheral lung lesions.Translational lung cancer research · 2026