INTRODUCTION: Sound can be divided into 2 forms of temporal information: envelope and fine structure. While cochlear implants (CIs) are effective in speech recognition in quiet, they are less effective for listening in noise and for music perception. This study aimed to compare a fine structure processing (FSP) strategy (FS4) with a previous coding strategy (HDCIS) in newly implanted CI users. METHODS: This prospective, randomized, cross-over, double-blind study included 19 participants between January 2017 and July 2018. Each participant was randomized to receive a map programmed with FS4 followed by a map programmed with HDCIS or vice versa. Tonal/vocal audiometry in quiet, music perception tests (Gabrielsson scale, interval test, melodic contour identification [MCI]), and quality of life (Hearing Implant Sound Quality Index [HISQUI]) assessments were conducted after 1 month of adaptation to each setting. RESULTS: FS4 significantly improved speech recognition in quiet (mean effect (ME) = 8.4%, 95% CI = 3.5-13.2%, p = 0.004, standardized effect size (SES) = 1.06), Gabrielsson scale scores (ME = 4.8, CI95% = 0.4-9.1, p = 0.047, SES = 0.67), and interval perception (ME = 18 Hz, 95% CI = 2.7-32.4 Hz, p = 0.024, SES = 0.78). No significant difference was found for MCI or HISQUI. A period effect was observed for speech recognition in quiet (p = 0.002), Gabrielsson scale (p = 0.005), interval test (p = 0.032), and HISQUI (p = 0.039). Eighty percent of the participants kept the FS4 setting after the study period. CONCLUSION: FS4 was associated with improved music perception and speech recognition in quiet compared to HDCIS in newly implanted CI users, suggesting potential advantages of FSP strategies. N° West V Ethics Committee: 2016-A01952-49, December 28, 2016.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers