The Establishment of National Diagnostic Reference Levels for Pediatric Chest X-Ray Examinations in Ethiopia.
Leul Adane Chemeda, Seife Teferi Dellie, Getu Ferenji, Daniel Zewdneh Solomon, Marema Jebessa Kumsa, Solomon Getachew
Department of Radiology College of Health Sciences Addis Ababa University Addis Ababa Ethiopia aau.edu.et. · Department of Internal Medicine St. Paul's Hospital Millennium Medical College Addis Ababa Ethiopia sphmmc.edu.et.
Key points
From the paper's conclusion- The 75th percentile of dose values determined in this study was higher than those published in the literature. The findings reveal significant variability in ESD across hospitals, primarily due to the use of inappropriate radiological parameters. A notable inconsistency in the selection of kV and mAs parameters was also observed, underscoring the need for standardized radiological protocols.
Abstract
BACKGROUND: It is well established that diagnostic X-ray practices must be optimized to keep patient radiation dose as low as possible with providing the diagnostic information required. The International Commission on Radiological Protection recommends diagnostic reference levels as a practical means to optimize radiation doses in diagnostic radiology and nuclear medicine. The aim of this study was to propose national diagnostic reference levels for routine pediatric chest X-ray examinations in Ethiopia. METHODS: Anthropometric and radiographic parametric data were collected from 975 patients. A standard patient was defined based on the mean values of weight, height, and BMI for the age groups of 5 to 10 and 10 to 15 years old. The exposure parameter, entrance surface air kerma, was calculated. NDRLs for entrance surface air kerma were defined as the 75th percentile of dose values. RESULTS: The suggested national diagnostic reference levels of chest X-ray examinations were 0.07 mGy and 0.53 mGy for age groups 5-< 10 and 10-< 15, respectively. CONCLUSION: The 75th percentile of dose values determined in this study was higher than those published in the literature. The findings reveal significant variability in ESD across hospitals, primarily due to the use of inappropriate radiological parameters. A notable inconsistency in the selection of kV and mAs parameters was also observed, underscoring the need for standardized radiological protocols.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Establishing diagnostic reference levels for pediatric chest radiography at a dedicated children's hospital in Hong Kong: A comparison of single criteria and hybrid grouping methods.Journal of medical imaging and radiation sciences · 2026
- Evolution of diagnostic reference levels in paediatric cardiac interventional procedures: a narrative review.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Determination of local diagnostic reference levels for mammography in Fars Province, Iran.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Optimization of pediatric head CT protocols and dose reduction by radiology technologists in Cameroon, Benin, and Senegal.Radiation protection dosimetry · 2026
- Local diagnostic reference levels for paediatric computed tomography based on European guidelines: anatomy- and indication-based approaches.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Comments on “Diagnostic reference levels based on clinical indications for paediatric computed tomography examinations at the Mohammed VI University Hospital Center in Marrakech-Morocco”Radioprotection · 2026