Undetected Intracellular Immunoglobulin Light Chains in Aberrant Plasma Cells of Multiple Myeloma by Monoclonal Antibody-Based Flow Cytometry: A Question Raised by a Single Case.
Jinyun Wang, Lulu Chen, Jie Xu, Yanchun Li, Ping Zhang
Department of Clinical Laboratory, Ninghai First Hospital, Ninghai, Zhejiang, 315600, China. · Department of Clinical Laboratory, Affiliated Hangzhou First People's Hospital, School of Medicine, Westlake University, Hangzhou, Zhejiang, 310006, China, westlake.edu.cn.
Key points
From the paper's conclusion- These findings were further supported by bone marrow morphology showing immature plasma cells and immunofixation electrophoresis confirming an IgG-λ monoclonal protein.
Abstract
This study presents a clinical case demonstrating the limitations of monoclonal antibodies (MAbs) in flow cytometric immunophenotyping for clonal disorders. Following standardized protocols, we performed surface antigen and intracellular light chain staining, using SSC/FSC gating with CD45/SSC and CD138/CD38 combinations to identify plasma cell populations. Flow cytometry analysis revealed abnormal plasma cell immunophenotypes (CD138+CD38+CD56+CD45-CD19- with partial CD117 expression) in a patient with multiple myeloma (MM), in which the MAb assay showed no detectable cytoplasmic κ or λ staining. Normal plasma cells showed characteristic CD138+CD38+CD45+CD19+CD56- expression with polyclonal light chain distribution (κ/λ ratio 1.43). These findings were further supported by bone marrow morphology showing immature plasma cells and immunofixation electrophoresis confirming an IgG-λ monoclonal protein. Our findings indicate a risk of false-negative results in intracellular light-chain detection using MAbs. Although polyclonal antibodies (PAbs) have been reported to potentially overcome the limitations of monoclonal reagents, unfortunately, no PAbs were available to test this in our case. Finally, this study systematically analyzes the reasons for the discordance between flow cytometric detection based on MAbs and other detection methods, and suggests that clonal assessment in MM requires integrating evidence from multiple detection methods for orthogonal validation to improve diagnostic accuracy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of a multiple myeloma synthetic cell mimic as an assay control for flow cytometry.Cytometry. Part B, Clinical cytometry · 2026
- Circulating tumor plasma cells quantification in newly diagnosed transplant ineligible real world multiple myeloma patients: Comparison of single-platform flow cytometry versus next generation flow.Cytometry. Part B, Clinical cytometry · 2026
- MyeloTrak: a sequencing-independent mass spectrometry assay for blood-based residual M-protein monitoring in multiple myelomaClinical Proteomics · 2026
- B-082 Evaluation of Minimal Residual Disease Detection Modalities in Multiple Myeloma: SPEP, Immunofixation, NGS, and Flow CytometryClinical Chemistry · 2026
- VS38c as a stable marker for identification of normal and malignant plasma cells.American journal of clinical pathology · 2026
- IGH and IGK Next-Generation Sequencing Testing in Multiple Myeloma: A Comparison of CD138-Positive Sorted and Unsorted Bone Marrow Samples.Annals of laboratory medicine · 2026