BioMed research international

Development of Polyclonal Anti-HBsAg IgY Antibodies and Comparative Evaluation of Glutaraldehyde Versus Periodate HRP Bioconjugation for Hepatitis B Diagnosis.

Felicienne Tchoufang Tchatchoua, Herman Philipe Nfombouot-Njitoyap, Yacouba Foupouapouognigni, Floriane Tekoutsap Manetsop, Arnaud Fondjo Kouam, Frédéric Nico Njayou, Simon Ngamli Fewou, Paul Fewou Moundipa

Faculty of Health Science, Université des Montagnes, Bangangté, Cameroon, udesmontagnes.org. · Laboratory of Pharmacology and Toxicology, Department of Biochemistry, Faculty of Science, University of Yaoundé 1, Yaoundé, Cameroon, uy1.uninet.cm.

Key points

From the paper's conclusion
  • We developed a polyclonal chicken IgY reagent against the immunodominant "a" determinant of HBsAg, using consensus peptides to limit immune escape: two 24-amino-acid consensus peptides from the major hydrophilic region were synthesised and used to immunise laying hens.

Abstract

Hepatitis B virus remains a major global health burden, particularly in sub-Saharan Africa, where affordable, specific and locally producible diagnostic tools are urgently needed. Conventional mammalian antibody-based assays can suffer from high background signals and interference in human sera. We developed a polyclonal chicken IgY reagent against the immunodominant "a" determinant of HBsAg, using consensus peptides to limit immune escape: two 24-amino-acid consensus peptides from the major hydrophilic region were synthesised and used to immunise laying hens. Antigen-specific IgY was extracted from egg yolks by salting-out precipitation, purified by peptide-affinity chromatography and conjugated to horseradish peroxidase by glutaraldehyde cross-linking and by sodium periodate oxidation. SDS-PAGE under reducing and non-reducing conditions confirmed the structural integrity of the purified IgY. Both chemistries yielded functional conjugates, but the periodate-oxidised conjugate gave a significantly higher signal across dilutions (Wilcoxon test, p = 0.016). In an adapted commercial sandwich ELISA applied to 10 HBsAg-positive and 10 HBsAg-negative sera, the in-house IgY-HRP conjugates produced optical density profiles that differed from the commercial biotinylated system (repeated-measures ANOVA, p < 0.0001) while separating positive from negative sera completely. Receiver-operating-characteristic analysis gave an area under the curve of 1.000 for all three detection systems, and at the Youden-derived cut-offs (OD450 0.77 for the glutaraldehyde and 0.80 for the periodate conjugate) both in-house reagents reached 100% sensitivity and 100% specificity (95% CI 72.2%-100% for each, n = 10 per group). This work establishes a low-cost, locally producible IgY-based platform for HBsAg detection, offering a scalable alternative for HBV screening in resource-limited, high-prevalence settings.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers