Evaluation of Below‐Knee Prosthetic Socket Mechanical Characteristics via Experimental and Numerical Methods
Saif M. Abbas, Sarah A. Hamood, Alaa Hussein Abdulaal
Key points
From the paper's conclusion- In summary, the prosthetic socket has acceptable structural integrity and user safety and function stability under physiological loading conditions.
Abstract
This study aimed at assessing the mechanical and biomechanical behavior of multilayer prosthetic sockets determined for transtibial amputation made up of polymer, carbon, and glass fibers. Measurements in one 45‐year‐old subject (83 kg, height: 167 cm) were evaluated, such as stump–socket interface pressure and strategy of composite mechanical properties. Two glass‐fiber layers increased the modulus of elasticity, yield strength, and ultimate tensile strength by 86.7%, 57.7%, and 32.2%, respectively, while two carbon‐fiber layers gave the following improvements of 51.85%, 89.3% and 64.3% as compared with group A (perlon only). The pressure test located on four sides (anterior, lateral, posterior, and medial) of the prosthetic socket. Peak pressure areas were located at the posterior (185 kPa) and lateral (160 kPa) aspects, highlighting the necessity to avoid anterior loading during score formation to protect the tibia and fibula. Results from static safety factor analysis were 0.85, 1.7, and 2.127 for Groups A (perlon), B (perlon and Glass fiber), and C (perlon and carbon fiber), respectively, with both Group B and SGR exceeding the commonly accepted biological safety limit. The maximum patellar tendon pressure was 125 kPa, and the lowest safety factor value of 2.8 in the anterodistal region showed that it had enough stability for being a structure. Close correspondence between finite element analysis (FEA) and experimental data (≤10% deviation) is indicative of the accuracy and validity of the computational model for comfort‐based design exploration. In summary, the prosthetic socket has acceptable structural integrity and user safety and function stability under physiological loading conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- An evaluated gait-driven finite element framework for transfemoral limb–socket interface biomechanics and socket design evaluationJournal of NeuroEngineering and Rehabilitation · 2026
- Mechanical evaluation of 3D-printed above-knee prosthetic sockets using a biofidelic hyperelastic silicone limb surrogate.Medical engineering & physics · 2026
- Personalized Prosthetic Socket with Embedded Multi-Location Pressure Sensing TechnologyTechnologies · 2026
- A Computational and Guided Strategy for PTB Transtibial Socket Design and Biomechanical Evaluation Using Finite Element AnalysisProsthesis · 2026
- Design and Bench Evaluation of a Sensor-Integrated Prosthetic-Leg Prototype.Sensors (Basel, Switzerland) · 2026
- Bioinspired auxetic metamaterial liners and sockets for transtibial prostheses: Energy absorption and stress redistributionJournal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · 2025