The European Physical Journal Plus

Demonstration of a cryogenic, switchable electron source for low-temperature detector calibration

D. A. Bennett, M. Borghesi, P. Campana, R. Carobene, A. Cattaneo, M. De Gerone, M. Faverzani, L. Ferrari Barusso,

Key points

From the paper's conclusion
  • We present the realization of a prototype of a compact, switchable electron source operating at cryogenic temperatures, demonstrated at energies ranging from 100 eV to 300 eV, and conceptually extendable to an arbitrary energy range.

Abstract

We present the realization of a prototype of a compact, switchable electron source operating at cryogenic temperatures, demonstrated at energies ranging from 100 eV to 300 eV, and conceptually extendable to an arbitrary energy range. The electrons are produced via photoelectric emission, induced by a LED illuminating two 400-nm-thick commercial aluminum layers inside a cryostat, and are subsequently accelerated by a voltage of up to 300 V. Detection is carried out with an array of transition-edge sensor (TES) microcalorimeters designed for X-ray detectors, where we successfully observe signals consistent with electrons produced at the source at rates of $$\gtrsim$$ ≳ 1 Hz and with an efficiency of $$\gtrsim 10^{-14}\ e^-/\gamma$$ ≳ 10 - 14 e - / γ .

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers