Validity of a Consumer-Grade Bioelectrical Impedance Device for Body Composition Assessment Compared with a Three-Compartment Model in Recreationally Active Adults.
Patrick G Saracino, Hannah K Eberhardt, Cecilia D Salas, Bernard Omolo
University of South Carolina Upstate, Human Performance and Health, Spartanburg, South Carolina, United States; SARACINP@uscupstate.edu. · University of South Carolina Upstate, Human Performance and Health, Spartanburg, South Carolina, United States.
Key points
From the paper's conclusion- This study aimed to validate a consumer multi-frequency bioelectrical impedance device against a 3-compartment (3C) model in recreationally active adults.
Abstract
This study aimed to validate a consumer multi-frequency bioelectrical impedance device against a 3-compartment (3C) model in recreationally active adults. Following an overnight fast and abstention from caffeine (12 h), alcohol (24 h), and strenuous exercise (24 h), participants underwent body composition assessment using a consumer multi-frequency bioelectrical impedance device (BIAH30; InBody Dial H30). A criterion 3C model was derived from body mass, body volume via air displacement plethysmography, and total body water via a validated multi-frequency bioelectrical impedance device. Fifty-five recreationally active participants (29M/26F; 22±4 years, 170.5±8.2 cm; 70.6±10.5 kg; 24.2±2.6 kg·m⁻²) completed the study. BF% (21.7±7.8% vs 19.9±7.8%) and FM (15.3±5.8 vs 14.0±5.8 kg) were higher with BIAH30 compared to 3C (p < 0.001), whereas FFM (55.5±10.6 vs 56.7±10.4 kg) was lower (p < 0.001). Bland-Altman analysis indicated mean biases of 1.9% (95% LOA: -4.0 to 7.7%), 1.3 kg (95% LOA: -2.7 to 5.3 kg), and -1.2 kg (95% LOA: -5.1 to 2.8 kg) for BF%, FM, and FFM, respectively, with no proportional bias. Lin's concordance correlation coefficient indicated moderate to strong agreement (CCC: 0.90-0.98). In conclusion, the InBody H30 demonstrated significant differences in BF%, FM, and FFM relative to 3C. Mean bias was small, no proportional bias was observed, and agreement was moderate to strong. However, 95% limits of agreement were wide, indicating notable individual variability. While the H30 may serve as a practical, low-cost option for group-level estimates or longitudinal tracking, it should not be considered interchangeable with multi-compartment models for precise individual-level measurement.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Determining body composition using different bioimpedance technologies: Is an agreement possible?Clinical Nutrition · 2025
- Body composition assessment: method comparison of DXA, InBody BIA and manual measurementsClinical Nutrition Open Science · 2026
- Threshold robustness of bioimpedance-derived adiposity classification across BMI categories in Chinese adults: a cross-sectional studyFrontiers in Nutrition · 2026
- Pairwise Agreement Among Field-Based Body-Fat Estimation Methods in Physically Active Adults: A Cross-Sectional Method-Comparison Study.Journal of functional morphology and kinesiology · 2026
- Agreement of Bioelectrical Impedance Analysis Prediction Equations with Dual-Energy X-Ray Absorptiometry for Tracking Resistance Training-Induced Changes in Fat-Free Mass in Older Women.Sports (Basel, Switzerland) · 2026
- The application of bioelectrical impedance analysis in assessing body composition of patients undergoing cardiac rehabilitationEuropean Heart Journal · 2026