Asymptomatic intrathecal baclofen pump catheter failure discovered during routine reservoir replacement: a case report.
Heeho Kim, Priyanka Singla
University of Virginia Department of Anesthesiology, Charlottesville, Virginia, USA sonokhh@gmail.com. · University of Virginia Department of Anesthesiology, Charlottesville, Virginia, USA.
Key points
From the paper's conclusion- Subjective clinical efficacy is an unreliable proxy for system patency. Aging hardware configurations require proactive radiographic monitoring and targeted surgical revision to minimize the risk of withdrawal.
Abstract
BACKGROUND: Intrathecal baclofen (ITB) therapy effectively manages severe spasticity. While mechanical complications typically cause therapy loss or withdrawal, structural disruptions can remain completely occult due to anatomical adaptations. This report explores the unique management of an incidentally discovered catheter transection during routine reservoir exchange. CASE PRESENTATION: A female patient in her 30s with cerebral palsy had been treated with ITB since 2011 without functional complaint, despite sluggish cerebrospinal fluid (CSF) flow noted at a 2019 pump exchange that did not prompt catheter revision. After relocating out of state, the patient returned in 2025 for a planned pump reservoir replacement, reporting no interval decline. Preoperative catheter port myelography was not successful due to failure of CSF aspiration from the catheter access port, and intraoperative exploration revealed the catheter was completely fractured at its fascial insertion point, despite the reported years of clinical stability. Fragment retrieval required partial laminectomy and multilayer dural repair; after new hardware placement and retitration, her therapeutic outcome improved dramatically. CONCLUSIONS: Subjective clinical efficacy is an unreliable proxy for system patency. Aging hardware configurations require proactive radiographic monitoring and targeted surgical revision to minimize the risk of withdrawal.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Management of Intrathecal Baclofen Pump System Malfunction and the Value of the Radio-Opaque Catheter Dye Study: A Retrospective Data Analysis of 49 Pump System Investigations Over a Seven-Year Period.Neuromodulation : journal of the International Neuromodulation Society · 2026
- Intrathecal Baclofen Pump Therapy for Medically Refractory Stiff Person Syndrome: A Multicenter Case Series.Neuromodulation : journal of the International Neuromodulation Society · 2026
- Intrathecal Therapy in a Patient With Solid Arthrodesis of the Interlaminar Spaces: A Case Report.A&A practice · 2026
- Early recognition and safe management of unintentional intrathecal placement of an epidural catheter during combined spinal-epidural labor analgesia: a case report in the context of the 2024 OAA guidelines.JA clinical reports · 2026
- Implanted intrathecal pump catheter migration rates in the management of cancer-related pain.Regional anesthesia and pain medicine · 2026
- Comparative pharmacotherapeutic efficacy and dose-related safety profiles of intrathecal baclofen: a systematic review and meta-analysis in neurological spasticity.Frontiers in pharmacology · 2026