科研速览 · Science Skim继续刷下去 · Keep skimming →
◆ Regional anesthesia and pain medicine2026-09-08

Complications leading to device explant in patients with chronic non-cancer pain receiving intrathecal drug delivery.

Salim Michel Hayek, Sameam Shahrestani, Jacob Blank, Beatriz Feijo, Jose Reyes, Phu Duong, Muskan Sidhu, Qiuchen Li, Ryan Shaner, Jennifer Murphy, Michael D Staudt

一句话结论 · In one sentence

Device-related infections are the leading cause for explants in IDD for chronic non-cancer pain. Explants due to loss of analgesic efficacy occurred at a lesser frequency with IDD compared with SCS explants in a similar patient population.

原始摘要(英文原文)· Original abstract
INTRODUCTION: Intrathecal drug delivery (IDD) is often used in the management of refractory chronic non-cancer pain, particularly persistent chronic low back pain in the setting of previous lumbar spine surgery (persistent spinal pain syndrome type 2 (PSPS2)). Complications of IDD leading to device explant in patients with chronic non-cancer pain have not been well characterized. This study evaluated complications leading to device explants in a large cohort of patients with chronic non-cancer pain treated with IDD. METHODS: A retrospective study was undertaken on patients with chronic non-cancer pain who underwent device implants by Pain Medicine physicians between June 2009 and December 2022 at a large tertiary center academic practice. Data collection included demographic and comorbidity data as well as the various indications for device explant. RESULTS: Data were collected from 215 patients with chronic non-cancer pain who underwent device implant and were followed for a mean duration of 80±48.8 months. The average age at the time of implant was 63.24±12.05 years, and the majority (184 patients) suffered from PSPS2. The explant rate was 26.1%, and device-related infection was the leading cause of explants occurring in 9.3% of patients and 5.9% of device surgeries. Loss of analgesic efficacy leading to device explant occurred in 5.6% of patients, but this was significantly less than the explant rate due to habituation noted with spinal cord stimulation (SCS) in a similar patient population. compared with other complications leading to device explants, infections had a much earlier time interval from device implant to explant. No demographic or comorbidity factors correlated with device explants. CONCLUSION: Device-related infections are the leading cause for explants in IDD for chronic non-cancer pain. Explants due to loss of analgesic efficacy occurred at a lesser frequency with IDD compared with SCS explants in a similar patient population.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Complications leading to device explant in patients with chronic non-cancer pain receiving intrathecal drug delivery. — 科研速览 Science Skim