BMJ connections. Surgery

Evaluation of devices used for endovascular repair of ruptured abdominal aortic aneurysms in the international consortium of vascular registries: an international, retrospective, multicentre, single-arm, open-label performance goal study.

Charles Banks, Kevin Mani, Xinyan Zheng, Jialin Mao, Salvatore Scali, Art Sedrakyan, Jack Cronenwett, Grace Wang,

Vascular Surgery and Endovascular Therapy, University of Alabama at Birmingham Health System, Birmingham, Alabama, USA. · Department of Surgical Sciences, Uppsala University, Uppsala, Sweden.

Trial · n=932

Key points

From the paper's conclusion
  • This proof-of-concept study shows harmonised international RWD can evaluate device performance in ruptured abdominal aortic aneurysm. Despite a comparator biased to favour OSR, EVAR met/exceeded the PG and reduced complications, supporting regulatory discussions to align instructions for use with guideline-endorsed EVAR-first strategies.

Abstract

IINTRODUCTION: The purpose of this study was to pilot a new concept for device appraisal-leveraging international real-world data (RWD) from the International Consortium of Vascular Registries (ICVRs) to assess aggregate endovascular aortic repair (EVAR) performance against a conservative performance goal (PG) derived from open surgical repair (OSR), with the intention of helping refine device labelling in collaboration with industry and the US Food and Drug Administration. RESEARCH DESIGN AND METHODS: International retrospective, multicentre, single-arm cohort, open-label clinical PG-based study within the ICVRs. Because OSR patients may not have been anatomically EVAR-eligible, the OSR cohort used for PG derivation was restricted to smaller diameters (<6.5 cm men; <5.5 cm women) to include patients eligible for EVAR. Analyses performed at Weill Cornell Medicine (MDEpiNet Center). RESULTS: OSR PG cohort: 932 patients (72±9 years average; 83.8% men) with 30% in-hospital mortality. EVAR cohort: 219 patients (75±10 years average, 83.1% men) with larger aneurysms (EVAR 7.3±2.0 cm vs OSR 5.3±1.3 cm; p<0.001). In-hospital mortality after EVAR was 18% (upper 95% CI 13% to 23%), exceeding the PG (p<0.001). EVAR showed lower dialysis (10.6% vs 15.7%; p=0.02), bowel ischaemia (1.8% vs 7.5%; p<0.001) and reoperation for bleeding (2.7% vs 7.1%; p=0.004). CONCLUSIONS: This proof-of-concept study shows harmonised international RWD can evaluate device performance in ruptured abdominal aortic aneurysm. Despite a comparator biased to favour OSR, EVAR met/exceeded the PG and reduced complications, supporting regulatory discussions to align instructions for use with guideline-endorsed EVAR-first strategies.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers