Comparison between choroidal thickness measurements from swept-source optical coherence tomography devices in highly myopic Chinese adults.
Syna Sreng, Huanhuan Tan, Kai Xiong Cheong, Yanfeng Jiang, Yvonne Ng, Valentina Bellemo, Xiangzhong Tang, Zhongping Chen,
Singapore Eye Research Institute, Singapore. · Changsha Aier Eye Hospital, Changsha, Hunan, China.
Key points
From the paper's conclusion- Vendor-provided software produced systematic differences between DREAM OCT and Triton OCT, whereas the deep learning model produced CT measurements with significantly reduced variability. When evaluating CT measurements from different platforms, deep learning-based approaches may provide more comparable results than vendor-provided software.
Abstract
BACKGROUND/AIMS: Reproducibility of choroidal thickness (CT) measurements is imperative when comparing choroidal imaging from different optical coherence tomography (OCT) platforms particularly in high myopia. This study compared CT measurements from commercial swept-source OCT devices in highly myopic Chinese adults using vendor-provided software and a deep learning segmentation model. METHODS: 97 participants underwent imaging with DREAM OCT and Triton OCT. CT was measured across the Early Treatment Diabetic Retinopathy Study (ETDRS) grid using vendor-provided software and a deep learning model, with ocular magnification correction. Agreement was evaluated using mean differences with 95% CIs, intraclass correlation coefficients (ICCs) and Bland-Altman plots. RESULTS: With vendor-provided software, DREAM OCT yielded thicker CT than Triton OCT (global ETDRS: 221.48±66.88 µm vs 156.90±57.64 µm), with mean differences of 61.04-71.22 µm (all p<0.001). Global agreement was moderate (ICC=0.63; 95% CI 0.56 to 0.68). Deep learning-derived CT measurements showed much smaller inter-device differences (DREAM: 173.47±59.54 µm; Triton: 171.21±56.62 µm), yielding a mean global difference of 2.26 µm with limits of agreement from -17.38 to 21.90 µm. Sectoral CT differences were similarly minimised (ICCs>0.90 across all ETDRS regions). The deep learning model demonstrated significantly superior inter-device agreement (global ICC=0.98 vs 0.63; p<0.05). CONCLUSIONS: Vendor-provided software produced systematic differences between DREAM OCT and Triton OCT, whereas the deep learning model produced CT measurements with significantly reduced variability. When evaluating CT measurements from different platforms, deep learning-based approaches may provide more comparable results than vendor-provided software.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Choriocapillaris Thinning in Myopic Macular Degeneration Revealed by Swept-Source Optical Coherence Tomography C-Scan Imaging.Bioengineering (Basel, Switzerland) · 2026
- Comparison of Biometric Measurements Between Anterior Segment Hyperparallel and Swept-Source Optical Coherence Tomography.Translational vision science & technology · 2026
- Associations of Retinal Curvature With Choroidal Thickness and OCTA-Derived Choroidal Flow-Density Metric in High Myopia: A Two-Center OCTA Study of Interocular AsymmetryTranslational Vision Science & Technology · 2026
- Agreement of corneal thickness measured by Scheimpflug, SS-OCT, and specular microscopy in Chinese myopia.BMC ophthalmology · 2026
- Beyond tessellation: AI-derived choroidal vessel density on multi-channel cSLO as a surrogate biomarker for OCT choroidal parameters.Frontiers in medicine · 2026
- Interpretable Automated Detection and Measurement of Dome-Shaped Macula on Optical Coherence TomographymedRxiv · 2026