Automated oxygen flow titration for infants with bronchiolitis: a multicentre randomised controlled trial
Pierrick Cros, Adélie Martin, Maëlys Consigny, Tiphaine Bihouee, Alexandra Masson, Lea Gaitan, Jean-Michel Roué
Centre Hospitalier Régional Universitaire de Brest · Centre Hospitalier Universitaire de Nantes
Key points
From the paper's conclusion- Conclusions While we did not find evidence of a reduced length of stay, which may be due to low study recruitment, we did find the FreeO2 device yielded a positive reduction in flow rates of oxygen.
Abstract
Objective To compare automated closed-loop oxygen therapy using the FreeO2 device versus conventional manual oxygen therapy with regard to length of hospital stay among spontaneously breathing infants with bronchiolitis. Design Multicentre randomised controlled trial conducted from 2018 to 2023. Setting Ten paediatric departments across France. Patients 103 infants, 1–12 months of age, hospitalised for acute bronchiolitis requiring oxygen therapy with SpO 2 <92%, excluding severe bronchiolitis. Interventions Patients were randomised to receive either automated oxygen administration via the FreeO2 device or conventional manual administration. Endpoints The primary endpoint was length of hospital stay in hours from emergency admission to discharge. Secondary endpoints included re-hospitalisation at 7 and 30 days, use of non-invasive ventilation (NIV), transfer to intensive care within 3 days and time spent in hypoxia (SpO2<92%) or hyperoxia (SpO 2 >98%). Results The median hospital stay was 71.00 hours (IQR 45.8–97.6) in the FreeO2 group and 69.6 hours (IQR 47.5–116.7) in the manual group (p=0.39). The percentage of time within the oxygen target zone was 89.4% in the FreeO2 group vs 74.9% in the manual group (p<0.05). Median oxygen flow was 0.1 L/min (IQR 0.1–0.2) in the FreeO2 group vs 0.3 L/min (IQR 0.2–0.5) in the manual group (p<0.05). The groups did not significantly differ in re-hospitalisation at D7 (p=1.00) or D30 (p=1.00) or NIV use (p=0.47). Conclusions While we did not find evidence of a reduced length of stay, which may be due to low study recruitment, we did find the FreeO2 device yielded a positive reduction in flow rates of oxygen.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- High‐Flow Oxygen Therapy vs. Continuous Positive Airway Pressure in Hospitalised Bronchiolitis: A Meta‐AnalysisActa Paediatrica · 2025
- Automatic versus manual control of oxygen and neonatal clinical outcomes in extremely preterm infants: a multicentre, parallel-group, randomised, controlled, superiority trialThe Lancet Child & Adolescent Health · 2026
- High-Flow Nasal Cannula and Virological Surveillance in Severe Bronchiolitis: A Retrospective PICU Study in Southern Brazil.Journal of medical virology · 2026
- Automated Control of Oxygen Therapy: A State-of-the-Art Review of Closed-Loop Devices, with a Rule-Based, Network-Independent System (EMILY) as a Working Exemplar.Seminars in respiratory and critical care medicine · 2026
- 23 Reducing high-flow nasal cannula overuse in bronchiolitis for high-value carePaediatrics & Child Health · 2026
- Nebulized Epinephrine Versus Placebo in Infants with Moderate-to-Severe Bronchiolitis: A Randomized Controlled Trial.Children (Basel, Switzerland) · 2026