Algorithm Comparison for Oxygen Saturation Estimation in Whole Blood Using Diffuse Reflectance Hyperspectral Imaging.
Michelle Bryarly, Minh Ha Tran, Arrsh Ali, Ling Ma, Baowei Fei
Department of Bioengineering, University of Texas at Dallas, Richardson, TX, USA.
Key points
From the paper's conclusion- We determined experimentally that sodium dithionite consistently deoxygenated whole blood by 35% for every 1 mg/mL of sodium dithionite to hemoglobin.
Abstract
Hyperspectral imaging (HSI) has been used in vivo to estimate the oxygenation level (SO2) of whole blood inside blood vessels. However, little literature is devoted to validating the estimation. The goal of this research is to compare the performance of three commonly used methods: the Kubelka-Munk algorithm, optical density algorithm, and inverse Monte Carlo method, for the task of estimating SO2 from diffuse reflectance spectra. Validation was performed using two data types: Monte Carlo simulated data and ex vivo sheep blood that was deoxygenated using sodium dithionite (Na2S2O4). The SO2 of sheep blood was confirmed using a blood gas analyzer. We measured the diffuse reflectance using two types of HSI cameras: a push-broom and a snapshot HSI camera system. Our results showed that visible wavelengths (400 - 600 nm) are preferred for micro-vessels with diameters less than 100 microns, and near-infrared wavelengths (600 - 900 nm) are preferred for larger vessels. We determined experimentally that sodium dithionite consistently deoxygenated whole blood by 35% for every 1 mg/mL of sodium dithionite to hemoglobin. It was found that the inverse Monte Carlo method performed well in estimating SO2, with an average estimation error of 5.53% and 6.97% for ex vivo sheep blood and Monte Carlo spectra, respectively. Our findings provide a basis for determining future methods for in vivo SO2 estimations, including surgical HSI applications that rely on accurate perfusion assessment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development and validation of a reflectance-based approach for oxygen saturation estimation using high-speed hyperspectral imaging.Proceedings of SPIE--the International Society for Optical Engineering · 2026
- Physiological Parameter Estimation from Reduced-Wavelength Hyperspectral Data: A Model-Based StudyCurrent Directions in Biomedical Engineering · 2026
- Frequency-domain dynamic speckle dual-wavelength StO2 imaging: Feasibility of occlusion localizationApplied Physics Letters · 2026
- Real-Time SpO2 Estimation from Chest Reflectance Photoplethysmography: Algorithm Development and Clinical Validation Against Arterial SaO2.Sensors (Basel, Switzerland) · 2026
- Performance validation of a new finger-root-adapted ring-shaped oximeter sensor during stepwise steady-state inspired hypoxia.Journal of clinical monitoring and computing · 2026
- A PPG sensor-based multi-feature machine learning approach for wide-range and high-precision SpO2measurement.Physiological measurement · 2026