DEVELOPMENT AND PRELIMINARY BENCHTOP EVALUATION OF A NOVEL ATHERMAL SEPTOSTOMY CATHETER.
Majid Jadidi, Jason MacTaggart, Bahman Kargarbahrkhazar, Alexey Kamenskiy
University of Nebraska Omaha, Omaha, NE. · University of Nebraska Medical Center, Omaha, NE.
Key points
From the paper's conclusion- These results validate the Aquablade's ability to perform precise, athermal septostomy without collateral damage.
Abstract
Chronic Type B aortic dissection (cTBAD) often requires aortic septostomy to depressurize the false lumen and enable effective endovascular repair. However, current "off-label" techniques, such as mechanical "cheese-wire" or electrosurgical cutting, lack precision and carry significant risks of uncontrolled tearing and collateral thermal injury. To address this unmet need, we developed the Aquablade catheter, which utilizes a dual-guidewire system and a contained, athermal, high-velocity fluid jet to precisely cut the aortic septum longitudinally. This study reports the design and preliminary ex vivo benchtop evaluation of the device. Functional prototypes were tested on human aorta and standard vascular graft materials in a custom benchtop setup. The device successfully created clean, continuous longitudinal cuts through both biological tissue and synthetic graft material. Visual inspection confirmed that the device's deflector anvil protected off target areas from injury. These results validate the Aquablade's ability to perform precise, athermal septostomy without collateral damage. This technology offers a promising, safer alternative for managing complex chronic dissections, supporting further evaluation in preclinical models.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A BENCHTOP PLATFORM FOR EVALUATING SEPTOSTOMY IN AORTIC DISSECTION.Proceedings of the ... Design of Medical Devices Conference. Design of Medical Devices Conference · 2026
- Transcatheter Electrosurgical Aortic Septotomy Across the Spectrum of Type B Aortic Dissection: Feasibility, Technical Lessons, and Bailout Strategies.EJVES vascular forum · 2026
- Integrated system combining cerebral protection and active steering directional puncture for thoracic aortic in situ fenestration.JVS-vascular science · 2026
- Use of a novel integrated dilator-needle system for trans-septal puncture in three dogs undergoing left atrial decompression.Journal of veterinary cardiology : the official journal of the European Society of Veterinary Cardiology · 2026
- Robust water-activated tissue adhesive patch for arterial/heart wound closure after intervention surgeryNature Communications · 2026
- Subcoronary Implantation of Decellularized Aortic Allografts-An Ex Vivo and In Vivo Feasibility Study.Annals of thoracic surgery short reports · 2026