Predicting the 1:10 Diluted MFI of Luminex Single Antigen Beads: A Tool for Assessing Antibody Strength and Desensitisation Eligibility.
Alexis Janin, Cédric Usureau, Marie-Joëlle Apithy, Nicolas Congy-Jolivet, Barbara Proust, Camille Humeau, Isabelle Jollet, Grégory Gatouillat,
EPFL, Lausanne, Switzerland. · Unité Mixte de Recherche (UMR) 1342, Institut de Recherche Saint-Louis (IRSL), Université Paris-Cité, Hôpital Saint-Louis, Paris, France.
Key points
From the paper's conclusion- We retrospectively analysed paired neat and 1:10 diluted MFI from two cohorts: a monocentric development cohort (Saint-Louis Hospital, AP-HP) used for model fitting, and a multicentric external validation cohort from five histocompatibility laboratories used to assess transportability.
Abstract
The Luminex Single Antigen (LSA) bead assay is the reference method for anti-HLA antibody detection in transplantation. Mean Fluorescence Intensity (MFI) is a semi-quantitative measure of antibody strength but saturates at high values. Serum dilution, recommended to refine quantification, has a non-linear and poorly characterised relationship to neat MFI. We retrospectively analysed paired neat and 1:10 diluted MFI from two cohorts: a monocentric development cohort (Saint-Louis Hospital, AP-HP) used for model fitting, and a multicentric external validation cohort from five histocompatibility laboratories used to assess transportability. The dilution function was modelled by locally weighted scatterplot smoothing (LOWESS) regression, with hyperparameters selected by the Bayesian Information Criterion (BIC); 95% prediction intervals were estimated locally from candidate residual distributions (Gaussian and SkewNormal). The empirical low-MFI slope was compared to the theoretical 0.10 expected from a 1:10 dilution. The development dataset comprised 237,547 paired measurements from 644 patients and 1,965 sera. The retained model was a degree-2 locally quadratic estimator with Gaussian kernel and bandwidth 1500; residuals were best fitted by a SkewNormal distribution. For a neat MFI of 8,000, the predicted diluted MFI was 2,130 (95% PI: 650-4,640). The empirical low-MFI slope was 0.219, substantially exceeding the theoretical 0.10. External validation on 18,642 measurements from 104 sera across four laboratories yielded a 95% prediction interval coverage of 93.9%. A non-parametric model with locally adapted prediction intervals is provided to predict the 1:10 diluted MFI without performing serum dilution, supporting clinical decision-making. Each predicted value comes with its prediction interval, whose width increases with neat MFI and is greatest in the saturating range.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Signal- and Bead-Dependent Inter-Operator Variability in Single-Center Luminex-Based HLA Antibody TestingJournal of the Oman Medical Association · 2026
- MARCo: An Interactive Web-Based Tool for Comprehensive Evaluation of HLA Antibody Reactivity in Luminex Single Antigen Bead Assays.HLA · 2026
- HLAResolver: A statistical framework for resolving HLA ambiguity in T-cell response databioRxiv · 2026
- Lipemia and serum index-related interference in Luminex-based anti-HLA antibody detection.Journal of immunological methods · 2026
- Advancing Quantitative ADA Detection Through Model Informed Assay Development (MIAD)The AAPS Journal · 2026
- Nanodisc-Based Platform for HLA Antibody Identification.Advanced functional materials · 2026