One-Year Outcomes of Pulmonary Vein Isolation With a Novel Cryoballoon in Asian Population.
Shinsuke Miyazaki, Yasuteru Yamauchi, Young Keun On, Masaomi Kimura, Kenji Ando, Keiichi Ashikaga, Osamu Inaba, Yasuaki Tanaka,
Institute of Science Tokyo, Tokyo, Japan. · Yokohama City Minato Red Cross Hospital, Kanagawa, Japan.
Key points
From the paper's conclusion- Background Cryoballoon ablations for pulmonary vein isolation (PVI) to treat atrial fibrillation (AF) are well established.
Abstract
BACKGROUND: Cryoballoon ablations for pulmonary vein isolation (PVI) to treat atrial fibrillation (AF) are well established. The POLAR SMART study was designed to collect prospective real-world data on acute and 12-month safety and effectiveness of the POLARx cryoballoon for de novo PVI. METHODS: POLAR SMART is a prospective, non-randomized, multicenter study (NCT05282823) with patients enrolled across 19 centers (Japan and South Korea) and followed up to 12 months. RESULTS: AF ablation was performed in 293 patients, 181 receiving only PVI. In the PVI-only cohort, acute PVI was achieved in 95.0% of patients, and 98.2% of pulmonary veins (PV) with the cryoballoon. Procedure time and fluoroscopy time were 75 ± 25 min and 18.5 ± 11.1 min, respectively. Time to isolation was 42 ± 32 s, with nadir temperature of -56.6 ± 6.9°C. The study met the primary safety endpoint with an event-free rate of 97.3%. The study effectiveness event-free rate at 12 months was 75.6% for entire cohort and 81.1% for PVI-only cohort (POLARx, n = 116; 79.3% and POLARx FIT, n = 65; 84.4%). AF recurrence in PVI-only cohort was 7.2% for POLARx and 6.2% for POLARx FIT. Ablation procedure increased heart rate (HR) significantly for entire cohort (p < 0.001). The optimal HR value at predischarge for 12-month outcome was calculated with 70 bpm (Predicted Probability 0.83). The effect of increased HR remained stable after 12 months for patients ablated with POLARx FIT. CONCLUSIONS: The POLARx and POLARx FIT cryoballoon systems are safe and effective for PVI in real-world use. POLARx FIT ablation was associated with a sustained HR increase over 12 months.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Acute efficacy and mid-term clinical outcomes of expandable vs. fixed-size cryoballoon ablation for paroxysmal atrial fibrillation: a two-center randomized studyEuropean Heart Journal · 2026
- Real-World Outcomes of Cryoballoon vs. Radiofrequency Ablation in Atrial Fibrillation: A Multicenter Propensity-Matched Analysis of Short- and Long-term EfficacyKorean Circulation Journal · 2026
- Safety, effectiveness and healthcare resource implications of an early discharge pathway for cryoballoon ablation: real-world evidence from China.Frontiers in cardiovascular medicine · 2026
- Comparative one-year outcomes of pulmonary vein isolation and roof line ablation using cryoballoon versus pulsed-field ablation in persistent atrial fibrillation with left atrial dilatationEuropean Heart Journal · 2026
- Pulmonary Vein Anatomy, Cooling Kinetics, and Recurrence After Cryoballoon Ablation for Atrial FibrillationJournal of Arrhythmia · 2026
- Proof of concept - retrospective virtual reality pulmonary veins modeling evaluating eligibility for atrial fibrillation cryoballoon ablationEuropean Heart Journal · 2026