Methods and Techniques for Closed-Loop Gastrointestinal Pacing.
Chaveeka Weerakoon, Leo K Cheng, Peng Du, Niranchan Paskaranandavadivel, Nipuni D Nagahawatte
Auckland Bioengineering Institute, University of Auckland, Auckland, New Zealand.
Key points
From the paper's conclusion- Closed-loop pacing significantly reduced pulse delivery, demonstrating the feasibility of a targeted, adaptive approach that may reduce pacing burden, although effects on motility and gastric emptying warrant further investigation.
Abstract
PURPOSE: Gastrointestinal slow waves are rhythmic electrical events that contribute to the coordination of motility. Functional motility disorders such as gastroparesis are linked to abnormal slow wave activity, and current treatments do not target the underlying electrical dysfunction. Gastrointestinal electrical pacing restores normal rhythms, but open-loop pacing lacks adaptation to underlying slow wave activity, leading to reduced energy efficiency and potential overstimulation. METHODS: A novel closed-loop pacing system was developed to detect gastric and intestinal slow waves and dynamically pace in response to the underlying rhythm. High-resolution electrode arrays were placed on the serosal surface of anesthetized pigs for simultaneous pacing and mapping. The control strategy comprised two steps: (i) continuous detection and classification of slow waves as normal rhythm (stomach: 15-25 s; small intestine: 3-5 s) or abnormal rhythm; and (ii) pacing delivery and entrainment assessment, with pulses (100-400 ms, 3-8 mA) triggered by abnormal rhythms or failed entrainment. RESULTS: Detection of slow wave events achieved an accuracy of 78% for both the stomach and the small intestine. Entrainment detection yielded an accuracy of 73% for the stomach and 67% for the small intestine. Closed-loop pacing successfully restored abnormal gastric and small intestinal slow wave periods to normal physiological ranges and reduced the number of pulses delivered by 60% compared to open-loop pacing. CONCLUSION: Closed-loop pacing significantly reduced pulse delivery, demonstrating the feasibility of a targeted, adaptive approach that may reduce pacing burden, although effects on motility and gastric emptying warrant further investigation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Endoscopic gastric pacing: A minimally invasive approach for slow wave entrainment in rhythmic and dysrhythmic states.American journal of physiology. Gastrointestinal and liver physiology · 2026
- Gastric slow-wave morphology in health and diseaseDigestive Diseases and Sciences · 2026
- Sedatives reduce human gastric slow-wave amplitude, assessed by a multicenter study of endoscopic gastric mappingmedRxiv · 2026
- Gastric electrophysiologyExpert Review of Gastroenterology & Hepatology · 2026
- High-frame-rate ultrasound analysis of gastric peristalsis using the peristaltic analyzer: a pilot feasibility studyFrontiers in Bioengineering and Biotechnology · 2025
- An Implantable In-Tube System for Quantitative Precision Vagus Neuromodulation and Feedback of Intestinal Motility.Advanced materials (Deerfield Beach, Fla.) · 2026