Journal of cardiovascular electrophysiology

Safety and Efficacy of Linear Ablation for Peri-Mitral Annular Atrial Flutter Using the Pentaspline Pulse Field Ablation Catheter.

Connor P Oates, Scott Grubman, William Whang, Mohit K Turagam, Daniel Musikantow, Joshua Lampert, Marc A Miller, Jacob S Koruth,

Mount Sinai Fuster Heart Hospital, Icahn School of Medicine at Mount Sinai, New York, New York, USA. · Department of Medicine, Icahn School of Medicine at Mount Sinai, New York, New York, USA.

Key points

From the paper's conclusion
  • While the pentaspline catheter can achieve acute procedural success when treating peri-mitral AFL with linear ablation of the mitral annulus, the long-term efficacy of this treatment strategy is modest.

Abstract

BACKGROUND: Despite widespread use of the pentaspline pulse field ablation (PFA) catheter to treat atrial fibrillation, there is limited data supporting its feasibility, safety, and effectiveness in creating linear lesions to treat peri-mitral atrial flutter (AFL). OBJECTIVE: To assess the safety and efficacy of using the pentaspline PFA catheter to treat peri-mitral AFL. METHODS: We performed a single-center retrospective review of consecutive patients receiving catheter ablation using the pentaspline PFA catheter to treat peri-mitral macroreentrant AFL. RESULTS: A total of 80 patients underwent catheter ablation for clinical peri-mitral AFL. Ablation was guided by electroanatomic mapping in all patients. AFL was treated with a lateral mitral line in 38 patients (47.5%), an anteroseptal mitral line in 31 patients (38.8%), a septal mitral line in 6 patients (7.5%), and an anterolateral mitral line in 5 patients (6.3%). Acute procedural success was achieved in 70 patients (87.5%) using the pentaspline catheter; the remaining 10 patients (12.5%) required additional radiofrequency ablation to achieve procedural success. Acute complications occurred in 4 (5.0%) patients: 2 (2.5%) patients sustained electrical isolation of the left atrial appendage, 1 (1.3%) had transient hematuria, and 1 (1.3%) had pericarditis. After a median follow-up of 322 days (IQR 192-432), 42 patients (52.5%) had recurrent AFL. Of these patients, 26 patients underwent a repeat ablation procedure-all demonstrated electrical breakthrough across the prior mitral line. CONCLUSIONS: While the pentaspline catheter can achieve acute procedural success when treating peri-mitral AFL with linear ablation of the mitral annulus, the long-term efficacy of this treatment strategy is modest.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Safety and Efficacy of Linear Ablation for Peri-Mitral Annular Atrial Flutter Using the Pentaspline Pulse Field Ablation Catheter — Science Skim