Epicardial Pulsed Field Ablation Across the Intact Pericardium: Feasibility, Safety Margins, and Tissue Selectivity in a Porcine Model.
Chin-Yu Lin, Ting-Yung Chang, Pei-Yi Wang, Yung-Nan Tsai, Shih-Lin Chang, Ding-Jun Juan, Cheng-Han Chan, Yuen Hoong Phang,
Division of Cardiac Electrophysiology, Cardiovascular Center, Taipei Veterans General Hospital, Taipei, Taiwan.
Key points
From the paper's conclusion- Epicardial PFA over an intact pericardium using a pentaspline catheter system could still create predictable lesions in the subepicardial area. The profound discrepancy between gross lesions and true necrosis suggests the pericardium acts as an electrical barrier. While massive overdoses cause non-selective damage, PFA's ability to create scars through a biological membrane inspires future catheter designs that could trap electrolytic microbubbles to eliminate embolic risks.
Abstract
BACKGROUND: Epicardial pulsed field ablation (PFA) holds promise, but in vivo safety, dose-responses, and the intact pericardium's barrier effect remain largely unexplored. METHODS AND RESULTS: Six pigs received 5, 10, 15, or 30 PFA shots via a pentaspline catheter (FARAWAVE system, 2.0 kV) placed externally on the intact left ventricular pericardium. Lesions were evaluated macroscopically and histologically at 1 month. PFA universally induced ventricular fibrillation requiring rescue shocks, and an observed numerical decline in left ventricular ejection fraction was noted at 1 month (72.6% to 50.1%). A distinct, nonlinear dose-response emerged. Standard doses (5-10 shots) yielded minimal true necrosis (0.15-2.63 cm3) despite broader gross lesions. At 15 shots, true necrosis expanded to 4.39 cm3, approaching the safety threshold. Conversely, an extreme 30-shot dose breached the pericardial impedance, causing a catastrophic spike in transmural necrosis (22.5 cm3) and early lethal heart failure. CONCLUSIONS: Epicardial PFA over an intact pericardium using a pentaspline catheter system could still create predictable lesions in the subepicardial area. The profound discrepancy between gross lesions and true necrosis suggests the pericardium acts as an electrical barrier. While massive overdoses cause non-selective damage, PFA's ability to create scars through a biological membrane inspires future catheter designs that could trap electrolytic microbubbles to eliminate embolic risks.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pulsed Field Ablation for Atrial Fibrillation: Mechanism, Clinical Evidence, and Emerging Applications.Cureus · 2026
- Complexity and Perplexity of Pulsed Field Ablation: An Engineering PerspectiveArrhythmia & Electrophysiology Review · 2025
- Pulsed field ablation: from concepts to technology.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2026
- Pulsed field ablation of atrial arrhythmias in patients with cardiac implantable electronic devices: a multicenter evaluation of electrical integrity and device safety.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Specific effects of Pulsed Field Ablation on the cardiac tissue, the nervous system and the conduction system in experimental and clinical studies.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2026
- Biophysical principles of pulsed field ablation.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2026