A 3D Print-Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps.
Aldo J Suria, Niccolò Crescenti, Benjamin Torner, Daniel Zimpfer, Marcus Granegger
Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.
Key points
From the paper's conclusion- Same NIH and pressure head SD range has shown the viability of the 3DP experimental framework, proposing an economic and rapid path for systematic haemocompatibility studies of parameters in RBPs.
Abstract
OBJECTIVE: Variability of 3D-printed rotodynamic blood pumps (RBP) indicators is a crucial factor in haemocompatibility evaluation. Thus, the aim is to investigate the reproducibility of a 3D printed-based framework for further use on systematic haemocompatible studies when changing key impeller parameters in RBPs. METHODS: Reproducibility was assessed in terms of standard deviation (SD), of the normalized index of haemolysis (NIH) and generated pressure head between 3 RBPs. A conventionally manufactured benchmark RBP, one 3D printed (3DP) exact replica, ( β 2 = 79 ° ), and one 3DP with a modification ( β 2 = 155 ° ) (n = 7). In vitro haemolysis experiments were carried out. Newly 3DP impellers were used for each experiment to study the manufacturing reproducibility. Moreover, pressure head-flow rate curve analysis and computational fluid dynamic (CFD) simulations were performed to support the findings. RESULTS: The mean NIH ± SD in mg/100 L for the 3 RBPs was 4.25 ± 1.95 for the benchmark, 6.40 ± 2.68 for the exact replica, and 7.75 ± 1.87 for the modified version. Only a significant difference (p < 0.05) between the benchmark and modified version was found. Conversely, no strong statistical differences were found between the 3DP RBPs. Furthermore, SD values of pressure heads were similar between all the groups. The resulting damage index for the 3DP geometries was 2.37 (·10-4) and 2.32 (·10-4) for the replica and modified version respectively. Suggesting a probable small effect when variating β 2 in free stream pumps. CONCLUSIONS: Same NIH and pressure head SD range has shown the viability of the 3DP experimental framework, proposing an economic and rapid path for systematic haemocompatibility studies of parameters in RBPs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative numerical and experimental evaluation of 3D-printed micromixers for biotechnological applications.Scientific reports · 2026
- Optimization of the Mechanical Response in MEX Additive Manufacturing of Thermoplastic Polyimide (PI): The Impact of Key Process Control SettingsACS Omega · 2025
- Fuzzy Logic-Based Modeling and Multi-Objective Optimization of Blood Pump Rotational Speed Considering Hydraulic and Hemolytic Performance.Medical engineering & physics · 2026
- CBCT in Dental Research: Is Complexity Necessary? An Exploratory Proof-of-Concept Study.Journal of functional biomaterials · 2026
- Machine-Learning-Assisted Quantitative Printability Assessment in Extrusion-Based Bioprinting-A Systems-Engineering Proof-of-Concept.Bioengineering (Basel, Switzerland) · 2026
- Taguchi-assisted Multi-Objective Differential Evolution for enhancing mechanical performance of 3D printed PETGMaterials Today Communications · 2026