Surrogate-Assisted Multi-Objective Optimization of an Axial-Flow Ventricular Assist Device Rotor Using CFD, Hybrid Surrogates, Sobol Analysis, and MOPSO.
Mohamed Bounouib, Reda Lakraimi, Hamza Isksioui, Mourad Taha-Janan, Wajih Maazouzi
Laboratory of Applied Mechanics and Technologies, ENSAM, Mohammed V University in Rabat, Rabat, Morocco. · Laboratory of Engineering Sciences and Applications (LSIA), National School of Applied Sciences (ENSAH), Abdelmalek Essaadi University, Al-Hoceima, Morocco.
Key points
From the paper's conclusion- The workflow supports design-space screening, trade-off identification, and targeted CFD confirmation of candidate geometries.
Abstract
BACKGROUND: This study presents a surrogate-assisted workflow for multi-objective optimization of an axial-flow ventricular assist device (VAD) rotor using 4809 steady-state CFD simulations spanning six geometric variables. METHODS: Repeated learning-curve analysis and blade-count-stratified Gaussian-process regression (GPR) assessed data efficiency and complemented tree-based surrogate models. Final selection retained GPR for pressure rise, torque, hemolysis index, platelet activation, and exposure time, and HistGradientBoosting for hydraulic efficiency and the low-shear fraction S < 50 Pa, with held-out R2 values of 0.968-1.000. Because hydraulic efficiency is derived from pressure rise and torque at the fixed operating condition, the primary MOPSO formulation used six objectives: maximizing pressure rise and S < 50 Pa while minimizing torque, hemolysis index, platelet activation, and exposure time; efficiency was retained as a secondary engineering metric. Ten independent MOPSO runs were pooled to characterize Pareto trade-offs, and five representative Pareto designs were re-evaluated by direct CFD. RESULTS: The balanced compromise used a 53.58° inlet blade angle, 20.93° outlet blade angle, three blades, a 0.222 mm clearance gap, 0.796 mm blade thickness, and a 20.00 mm rotor length. For the balanced compromise, mean surrogate-to-CFD relative error was 4.0%, with all individual errors below 10%. Relative to baseline, direct CFD confirmed reductions of 17.2% in torque, 55.6% in hemolysis index, 46.9% in platelet activation, and 27.0% in exposure time, together with a 4.5% increase in S < 50 Pa; pressure rise decreased by 22.4% and hydraulic efficiency changed by +0.7%. CONCLUSIONS: The workflow supports design-space screening, trade-off identification, and targeted CFD confirmation of candidate geometries.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Surrogate-assisted multi-objective optimization of an axial-flow ventricular assist device rotor using a large CFD-derived design space.Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine · 2026
- Tolerance-aware robust ranking of ventricular assist device rotor designs using strict feasibility and percentile-based criteria.The International journal of artificial organs · 2026
- Hybrid Blade Parameterization for Surrogate-Based Multi-Objective Optimization of an Axial Ring FanApplied Sciences · 2026
- Multi-objective optimization design of linear oscillating actuator based on hybrid surrogate modelScientific Reports · 2026
- Closed-Loop CFD-Driven Multi-Objective Bayesian Optimization of an Opposing Jet in Supersonic Flow for a Blunt BodyAerospace · 2026
- Multi-Objective Optimization of a Micro High-Speed Centrifugal Pump Based on an Adaptive Heterogeneous Ensemble Surrogate Model and NSGA-IIJournal of Physics: Conference Series · 2026