A Discrete Hemodynamic Control System: Proof-of-Concept In-Vivo Study in Canine Heart Failure Using a Pulmonary Artery Catheter.
Yasuyuki Kataoka, Kazunori Uemura, Takuya Nishikawa, Yukiko Fukuda, Jon Peterson, Kei Sato, Hiroki Matsushita, Hidetaka Morita,
NTT Research, Inc. Sunnyvale CA 94085 USA. · National Cerebral and Cardiovascular Center Suita 564-8565 Japan.
Key points
From the paper's conclusion- Conclusions: This proof-of-concept in-vivo study demonstrates the feasibility of implementing discrete two-degree-of-freedom hemodynamic control under pulmonary artery catheter-based monitoring and phlebotomy-mediated volume removal in anesthetized canine models of tachycardia-pacing-induced heart failure.
Abstract
Objective: Managing heart failure requires simultaneous regulation of mean arterial pressure, cardiac output, and left atrial pressure, which is challenging due to complex cardiovascular interactions and limited monitoring. This study presents a proof-of-concept in-vivo evaluation of a discrete two-degree-of-freedom hemodynamic control system designed to regulate these three hemodynamic variables through titration of cardiovascular interventions under pulmonary artery catheter monitoring. Results: Across five consecutive trials in three canine models of heart failure induced by tachycardia pacing, the cohort-averaged mean arterial pressure, cardiac output, and left atrial pressure were each within ±10% of their respective targets when evaluated using the final three intervention measurements, with absolute percentage errors decreasing from 16.8% to 9.0%, 23.4% to 3.8%, and 87.5% to 5.5%, respectively. In a more severely complicated heart failure model with coronary embolization, outcomes revealed control limitation primarily due to pharmacological resistance, highlighting the importance of choosing the appropriate target value. Conclusions: This proof-of-concept in-vivo study demonstrates the feasibility of implementing discrete two-degree-of-freedom hemodynamic control under pulmonary artery catheter-based monitoring and phlebotomy-mediated volume removal in anesthetized canine models of tachycardia-pacing-induced heart failure. Further studies are required to evaluate the effects of inter-subject variability and clinical generalizability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Automatic control of arterial blood pressure: computer, benchtop, and animal models.Frontiers in medical technology · 2026
- Feasibility of pacing with digital twins of brainstem circuits to restore heart rate variability in heart failureEuropean Heart Journal · 2026
- Hydraulic performance of 14-G hemodialysis catheters versus an 8-Fr arterial cannula in a mock 3-kg small-dog cardiopulmonary bypass circuit.PloS one · 2026
- From Hemodynamics to Energetics: Parameter Identification in a Closed-Loop Model of the Cardiovascular System.International journal for numerical methods in biomedical engineering · 2026
- Presentation of preload and afterload to isolated heart preparationsThe Journal of Physiology · 2026
- Invasive pulmonary artery pressure monitoring in heart failure: evidence, clinical application, and future directions.Heart failure reviews · 2026