Reducing RF-Induced Heating of DBS in 3 T MRI Using Dual-Role Receive Arrays as Wireless Resonators: A Simulation Study.
Zhonghao Zhang, Ming Lu, Zhengyi Lu, Yuankai Huo, Xinqiang Yan
Department of Electrical and Computer Engineering, Vanderbilt University, Nashville, TN 37235, USA. · Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, TN 37232, USA.
Key points
From the paper's conclusion- The dual-role coil demonstrated a high degree of flexibility in controlling the transmit field and reducing RF-induced heating at DBS implants, offering a novel approach to mitigate RF-induced heating of the implants in MRI.
Abstract
Magnetic resonance imaging (MRI) scanning remains largely restricted to specific modalities, typically involving low radiofrequency (RF) power levels and stringent protocols for patients with deep brain stimulation (DBS) implants, due to safety concerns related to RF-induced heating of the implants. A 6-channel dual-role head coil array capable of modulating the electric-field (E-field) distribution was designed and evaluated using electromagnetic (EM) simulations. By optimizing the resonant frequency of each coil element during RF transmission, the transmit field was reshaped, leading to a significant reduction in RF-induced heating near the DBS lead tip. The proposed method was validated across two scenarios of increasing complexity: 1) a simple straight conductive wire for concept validation and 2) four realistic DBS leads representing complex real-world scenarios. The coil settings can be optimized either to suppress the E-field at a specific location, such as the DBS lead tip, or to suppress the peak specific absorption rate (SAR) across the entire human head. For location-specific E-field suppression, the simplified predefined-state control scheme and the fine-tuning genetic algorithm (GA)-based framework were implemented, achieving E-field reductions of 45.9% and 68.3%, respectively. For whole-head peak SAR suppression, the annealed Log-Sum-Exp (LSE)-Adaptive Moment Estimation (Adam) framework (LSE-Adam) was implemented, achieving an average 1 g SAR reduction of 72.09% across four realistic DBS lead models. The dual-role coil demonstrated a high degree of flexibility in controlling the transmit field and reducing RF-induced heating at DBS implants, offering a novel approach to mitigate RF-induced heating of the implants in MRI.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Head-only MRI of Deep Brain Stimulation devices at 0.5 T: Patient-derived approaches for estimating radiofrequency heating.NeuroImage · 2026
- A Guidewire Trajectory Generation Tool for Interventional MRI SimulationsarXiv · 2026
- Detailed Consideration of a Novel Meandered Dipole Array for Magnetic Resonance Imaging of the Head at 3 Tesla with Low Radiofrequency Power DepositionSensors · 2026
- RF safety assessment of dental retainers for human head MRI array coils at 9.4 T.Magnetic resonance imaging · 2026
- RF-EMF response modeling of implantable medical electrodes under 1.5T and 3T MRI exposure using a lumped element model approach.Physics in medicine and biology · 2026
- Impact of transducer position on B1 + homogeneity and RF safety in transcranial MR-guided focused ultrasound: An electromagnetic simulation study.Medical physics · 2026