Quantifying resolution in pink-beam dark-field X-ray microscopy: experiments and simulations.
M La Bella, H F Poulsen, S Staeck, N A Henningsson, M P Kabukcuoglu, C Detlefs, C Yildirim
Technical University of Denmark, 2800 Kgs Lyngby, Denmark. · Institute for Photon Science and Synchrotron Radiation (IPS), Karlsruhe Institute of Technology (KIT), 76344 Eggenstein-Leopoldshafen, Germany.
Key points
From the paper's conclusion- Pink-beam dark-field X-ray microscopy (pDFXM) is a powerful emerging technique for time-resolved studies of microstructure and strain evolution in bulk crystalline materials.
Abstract
Pink-beam dark-field X-ray microscopy (pDFXM) is a powerful emerging technique for time-resolved studies of microstructure and strain evolution in bulk crystalline materials. In this work, we systematically assess the performance of pDFXM relative to monochromatic DFXM when using a compound refractive lens as the objective. Analytical expressions for the spatial and angular resolution are derived and compared with numerical simulations based on geometrical optics and experimental data. The pink-beam configuration provides an increased diffraction intensity depending on the deformation state of the sample, accompanied by a general tenfold degradation in angular resolution along the rocking and axial strain (longitudinal) directions. This trade-off is disadvantageous for axial strain mapping but can be advantageous in cases where integrated intensities are needed. For a perfect crystal under parallel illumination with a pink beam, our results show that chromatic aberration is absent, whereas under condensed illumination it becomes significant. The aberration is shown to depend strongly on the local distortion of the crystal. Weak-beam imaging conditions, such as those required for resolving dislocations, are shown to remain feasible under pink-beam operation and may even provide an improved signal-to-noise ratio. The higher incident flux, enhanced by nearly two orders of magnitude, results in a ∼30× increase in diffracted intensity; beam heating effects are quantified and implications for optimized scanning protocols are discussed.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative analysis of parametric X-ray radiation from diamond powders with varying particle sizesNuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms · 2026
- Diamond compound refractive lenses for high-energy dark field X-ray microscopy.Journal of synchrotron radiation · 2026
- Multilayer Laue lenses for enhanced spatial resolution in dark-field x-ray microscopyApplied Physics Letters · 2026
- Bridging grain mapping and dark field X-ray microscopy for multiscale diffraction imaging.Scientific reports · 2026
- Quantitative polarimetric characterization of micron-scale scratches on transparent surfaces using Scheimpflug dark-field imagingMeasurement Science and Technology · 2026
- Pixel-wise quantitative dark-field imaging in a lab-based dual-phase X-ray grating interferometer.Optics express · 2026