Zero-Dead-Time Strontium Lattice Clock with a Stability at 10^{-19} Level.
Xiao-Yong Liu, Peng Liu, Jie Li, Yu-Chen Zhang, Yuan-Bo Wang, Zhi-Peng Jia, Xiang Zhang, Xian-Qing Zhu,
Hefei National Research Center for Physical Sciences at the Microscale and School of Physical Sciences, University of Science and Technology of China, Hefei 230026, China. · University of Science and Technology of China, Shanghai Research Center for Quantum Sciences and CAS Center for Excellence in Quantum Information and Quantum Physics, Shanghai 201315, China.
Key points
From the paper's conclusion- These results represent a more than ninefold improvement over a conventional single-ensemble clock, highlighting its potential for next-generation timekeeping applications.
Abstract
Optical atomic clocks play a crucial role in fundamental physics, relativistic geodesy, and the future redefinition of the Systeme International second. Standard operation relies on cyclic interrogation sequences, which alternate between atomic interrogation and dead time used for state preparation and readout. This approach introduces the Dick effect, where laser frequency noise aliases onto the atomic transition frequency. Although reducing laser noise improves clock stability, the Dick effect remains a key limitation. In this Letter, we demonstrate a zero-dead-time optical clock based on two interleaved ensembles of cold ^{87}Sr atoms. Our system significantly suppresses this noise and achieves a fractional frequency instability at the 10^{-19} level between 10 000 and 20 000 s over repeated measurements, with a best value of 2.9×10^{-19} at τ=20000 s. The estimated long-term stability based on the combined data of these measurements reaches 2.5×10^{-19} at 1 day. These results represent a more than ninefold improvement over a conventional single-ensemble clock, highlighting its potential for next-generation timekeeping applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Spatially defined Rabi spectroscopy for uninterrupted optical clock interrogation.Nature communications · 2026
- Compact two-photon optical clock with a long-term stability of 8.1 × 10−15 without linear drift removalReview of Scientific Instruments · 2025
- Improved systematic evaluation of a strontium optical clock with uncertainty below 1 × 10 − 18Metrologia · 2026
- Clock Precision beyond the Standard Quantum Limit at 10^{-18} Level.Physical review letters · 2025
- High-stability offset-frequency locking of two lasers using a balanced filter discriminatorApplied Physics Letters · 2026