bioRxiv : the preprint server for biologyOpen accessbioinformatics

A community audit of error control in crosslinking mass spectrometry.

Lutz Fischer, Bruno C Amaral, Micha Johannes Birklbauer, James E Bruce, Paulo C Carvalho, Robert J Chalkley, Milan A Clasen, Viktoria Dorfer,

Technische Universität Berlin, Chair of Bioanalytics, 10623 Berlin, Germany. · Department of Biochemistry and Molecular Biology, University of Calgary, Calgary, AB, Canada.

Key points

From the paper's conclusion
  • In a community effort, we audited error control in twelve crosslinking mass spectrometry pipelines using two ground-truth designs that expose a lower bound on false identifications.

Abstract

In a community effort, we audited error control in twelve crosslinking mass spectrometry pipelines using two ground-truth designs that expose a lower bound on false identifications. Some pipelines passed, whereas for others these verifiably false identifications alone exceeded the requested false discovery rate up to 29-fold at the protein-protein interaction level. The audit also tracked improvement between software versions, supporting its use for regression testing.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.