Gravity-driven flow for intracranial saccular aneurysm filling.
Tirosh Mekler, Maria Khoury, Yevgeniy Kreinin, Rami Fishler, Nir Natan, Ron Bessler, Alexander Leshansky, Josué Sznitman,
Department of Biomedical Engineering, Technion-Israel Institute of Technology , Haifa, Israel. · Department of Chemical Engineering, Technion-Israel Institute of Technology , Haifa, Israel.
Key points
From the paper's conclusion- Our findings show that slow, gravity-assisted injections minimize bypass and spillover, enabling targeted aneurysm filling even in complex geometries.
Abstract
Intracranial aneurysms (IAs) are life-threatening cerebrovascular malformations in which complete and durable sac closure is the primary goal of endovascular treatments, including coiling, stent-assisted coiling and flow-diverter implantation. However, standard catheter-based approaches may fail to fully occlude the sac and are limited by the need for direct access. Liquid embolic agents offer better geometric conformity, but their deployment remains constrained by similar access issues. In this study, we present a gravity-driven approach for distal filling of IAs. Through a combination of analytical modelling, in vitro experiments with idealized aneurysm models, and patient-specific computational fluid dynamics (CFD) simulations, we demonstrate that efficient filling of aneurysm cavities from relatively remote locations may be achieved using a heavier injectable fluid, while properly adjusting tilt angle and injection flow rate. A phase diagram of flow regimes highlights the interplay between gravitational, inertial and viscous forces critical to successful filling. Our findings show that slow, gravity-assisted injections minimize bypass and spillover, enabling targeted aneurysm filling even in complex geometries. CFD simulations highlight the need for patient-specific planning of such a procedure accounting for personal vascular anatomy. This strategy may provide a safer and less invasive alternative to current endovascular techniques, especially for anatomically inaccessible aneurysms.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Patient-Specific Hemodynamic and Angiographic Evaluation of a Novel Flow-Diverting Device for Intracranial Aneurysms.Bioengineering (Basel, Switzerland) · 2026
- From Device Modeling to Blood Flow Simulation: A Study of the Woven EndoBridge.International journal for numerical methods in biomedical engineering · 2026
- Methodological Variability and Reporting Practices in Contemporary Patient-Specific Intracranial Aneurysm CFD Studies: A ReviewFluids · 2026
- Hemodynamic-Structural Multi-Fidelity Assessment of Endovascular Flow Diversion Strategies for Cerebral Aneurysm Treatment: An Integrated Computational and Experimental FrameworkbioRxiv · 2026
- Multicenter experience of pipeline embolization device used in small caliber vessels (< 2 mm) for intracranial aneurysm treatment and mid-term resultsFrontiers in Neurology · 2026
- Flow-driven magnetic microcatheter for superselective arterial embolizationScience Robotics · 2025