A Method to Reduce the Factors Affecting Iodine Enhancement for Computed Tomography Contrast Optimization.
Zahra Alyani Nezhad, Giuseppe V Toia, Sean D Rose, Rachel M Bladorn, Carrie M Bartels, Timothy P Szczykutowicz
Department of Medical Physics, University of Wisconsin-Madison, Madison, WI. · Henry Ford Hospital, Detroit, MI.
Key points
From the paper's conclusion- This study demonstrated a method to normalize effects of tube potential and beam hardening on CT number, yielding a metric proportional to iodine concentration.
Abstract
PURPOSE: Iodine contrast enhancement is assessed using computed tomography (CT) number, though this value is not proportional to iodine concentration. This study develops a metric for iodine enhancement that is directly proportional to iodine concentration. MATERIALS AND METHODS: Using studies previously reported in the literature, we parameterized the effect to create a correction for tube potential and patient-size-specific beam hardening on iodine CT number decrease. To validate the correction, we performed a study using different iodine concentrations, phantom sizes, and tube potentials. We also applied the correction to patient pulmonary angiogram data. Patients were grouped by tube potential [80, 100, 120, 140] (kV). Quantitative iodine contrast enhancement (Q-ICE) (ie, the difference between noncontrast and contrast-enhanced CT number) of the pulmonary artery (PA) was calculated. Pearson correlation coefficients are reported to characterize trends in PA Q-ICE across weight before and after correction. RESULTS: The range of CT numbers measured for [10, 15] mg/mL iodine inserts was [290, 433] HU before correction, and [24, 36] HU after correction. Before correction, median PA Q-ICE enhancement was [662, 490, 263, 175] HU for [80, 100, 120, 140] kV, with a range of 487 HU observed across all measurements. After correction, the median PA Q-ICE enhancement was [366, 373, 285, 249] HU, with a range of 124 HU. Pearson correlation coefficients significantly (P < 0.05) decreased after correction from -0.67 to -0.37. CONCLUSIONS: This study demonstrated a method to normalize effects of tube potential and beam hardening on CT number, yielding a metric proportional to iodine concentration.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Reduced contrast volume in coronary CT angiography: effects on coronary attenuationEgyptian Journal of Radiology and Nuclear Medicine · 2026
- Dose optimization in contrast-enhanced CT brain imaging using low tube voltage and reduced iodine concentration: a comparative study.BMC medical imaging · 2026
- Photon-counting CT iodine maps for late iodine enhancement: a retrospective feasibility study of image quality and reader confidence.European radiology experimental · 2026
- Evaluation of iodine detection and quantification limits in photon-counting CT-a phantom study.European radiology experimental · 2026
- Photon-counting detector CT for CT pulmonary angiography: a systematic review and meta-analysis.European radiology · 2026
- Accurate iodine quantification and residual error reduction with principal component analysis multimaterial decomposition using spectral CTMedical Physics · 2026