The Cost of Inaction: Ten Years of Neuraxial Misconnection Tragedies, 2014 to 2024.
Stacie Ethington, Amanda Volpe, Nina Smith, Kelsey Koym, Sonya Fogg, Debora Simmons
Institute for Safe Medication Practices, Fort Washington, PA. · Hartford HealthCare, Hartford, CT.
Key points
From the paper's conclusion- The search identified 95 records that detailed 110 cases of neuraxial misconnection.
Abstract
A tubing misconnection occurs when a delivery system, such as an intravenous line, tube, syringe, or catheter, is accidentally connected to a physiologically incompatible system. Misconnections can result in patient harm and death and have been reported for decades. The high-leverage risk mitigation strategy of using nonuniversal small-bore connectors to prevent connection to another route, system, or device is available. However, a recent survey conducted by clinical advisors working in coordination with the Global Engineered Device Supplier Association (GEDSA) found significant persistence of misconnection events. This paper details the methods and results of a systematic search and review of more recent (2014 to 2024) peer-reviewed literature, grey literature, and news stories to catalog cases of neuraxial misconnections. The search identified 95 records that detailed 110 cases of neuraxial misconnection. Twenty-three (20.9%) resulted in death, and 72 (65.5%) resulted in patient harm. Only 9 (8.2%) did not result in harm. A substantial number of cases occurred in the perioperative setting and, in particular, during labor and delivery. There were 6 medications associated with patient death. These medications were also associated with substantial patient harm. By providing information about the setting, type of error, medications involved, and level of harm, this paper describes misconnection events, explores contributing factors to the medical error, and provides proposed solutions to prevent misconnections. The ISMP Hierarchy of Effectiveness of Risk-Reduction Strategies is used to present layered and leveraged strategies to minimize the risk of these events reaching a patient.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Neuraxial Medication Administration Errors During Anaesthesia and Perioperative Management, Consequences, Contributing Human and Systemic Factors and Their Prevention: A Systematic Review.Pain and therapy · 2026
- When left isn't right: A case series of wrong-sided nerve blocks reported to webAIRS.Anaesthesia and intensive care · 2026
- Fatal accidental oxygen insufflation via suprapubic catheter.Forensic science, medicine, and pathology · 2026
- Inadvertent Epidural Administration of Diluted Unfractionated Heparin During Combined Spinal-Epidural Anesthesia: A Case Report and Patient Safety Analysis.Local and regional anesthesia · 2026
- Nursing Mechanisms of Intravenous Medication Loss in Adult Intensive Care: A Mixed‐Methods Systematic ReviewNursing in Critical Care · 2026
- When the interface fails: a patient-safety hazard hidden by the assumption of human error.Expert review of medical devices · 2026