Effect of Skin Pigmentation and Implantation Depth on Fluorescence-Based Oxygen Sensing in Ex Vivo Human Skin.
Mohamed M Ibrahim, Samer Kareem, Ahmed Elhessy, Yvonne M Rasko, Marc A M Mureau
. Department of Medicine, University of Maryland School of Medicine, Baltimore, MD, USA. · . Department of Medicine, University at Buffalo, NY, USA.
Key points
From the paper's conclusion- Fluorescence lifetime sensing demonstrated stable and consistent performance across varying skin pigmentation and implantation depths under controlled ex vivo conditions. These findings support its potential as a pigmentation-independent sensing approach and provide a foundation for further in vivo validation. Future research in physiological and clinical models is needed to confirm performance under dynamic conditions.
Abstract
BACKGROUND: Tissue oxygenation is crucial in reconstructive surgery, particularly when assessing flap viability. Current optical monitoring technologies, such as near-infrared spectroscopy (NIRS) and indocyanine green (ICG) dyes, encounter challenges due to variable signal strength influenced by skin pigmentation and tissue depth. Oxygen monitoring methods based on fluorescence lifetime could provide a more reliable and consistent alternative. METHODS: Benzoporphyrin-based oxygen sensors incorporated into polyHEMA hydrogel were assessed in ex vivo human skin representing all six Fitzpatrick skin types. We measured fluorescence intensity and lifetime at both intradermal and subcutaneous depths with controlled oxygen concentrations (100% and 12%). Additionally, sensor performance was tested across physiological ranges of temperature and pH to evaluate environmental robustness. RESULTS: We demonstrated that fluorescence lifetime measurements were consistent throughout all Fitzpatrick skin types at both intradermal and subcutaneous depths, while showing minimal sensitivity to pigmentation or implantation depth. In contrast, the absolute fluorescence intensity decreased with more skin pigmentation and demonstrated variability. Tissue oxygen tension (TOT) derived from lifetime measurements has significantly decreased under hypoxic conditions (12% oxygen) for all Fitzpatrick skin types, which confirms dependable oxygen responsiveness. Additionally, sensor performance remained consistent and stable within physiological temperature and pH ranges. CONCLUSIONS: Fluorescence lifetime sensing demonstrated stable and consistent performance across varying skin pigmentation and implantation depths under controlled ex vivo conditions. These findings support its potential as a pigmentation-independent sensing approach and provide a foundation for further in vivo validation. Future research in physiological and clinical models is needed to confirm performance under dynamic conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Anatomical and spectral foundation of skin pigmentation bias in optical oximetrymedRxiv · 2026
- Development and Assessment of Poly(ethylene glycol) Diacrylate-Embedded Metalloporphyrin for Phosphorescence Lifetime-Based In Vivo Monitoring of Oxygen Tension.ACS sensors · 2026
- Feasibility study of early pressure injury detection using single-chamber and dual-cavity optical sensor.Technology and health care : official journal of the European Society for Engineering and Medicine · 2026
- Phasor-based fluorescence lifetime imaging microscopy for label-free assessment of skin tissue responses to UV and 5G millimetre-wave exposureJournal of Optics · 2026
- Performance validation of a new finger-root-adapted ring-shaped oximeter sensor during stepwise steady-state inspired hypoxia.Journal of clinical monitoring and computing · 2026
- Biocompatible fluorescein-based temperature sensor for applications in endoscopic surgery.Methods and applications in fluorescence · 2026