Agreement Between Medical Laboratory Technologists and Pathologists' Assessment of ERBB2 (HER2) Fluorescence In Situ Hybridization Across Breast and Gastric Scoring Criteria.
Cameron Beech, Julia Bridge, Anna Matynia, Valarie McMurtry, Kajsa Affolter, Daniel J Albertson, Parisa Adelhardt, JongTaek Kim,
Departments of Pathology, University of Utah. · Institute for Experimental Pathology, ARUP Laboratories, Salt Lake City, UT.
Key points
From the paper's conclusion- In summary, distinct HER2 FISH technologists HER2/CEP17 ratio and HER2 signal number per cell values were identified where concordance with pathologists results exceeded 99%, suggesting cases in these ranges may require only limited pathologist slide review, potentially allowing for more efficient triaging and sign-out of HER2 FISH.
Abstract
ERBB2 (HER2) fluorescence in situ hybridization (FISH) guides therapy selection for solid tumors, including breast and gastric cancers. While laboratory technologists typically perform the initial enumeration of HER2 FISH, laboratories differ in the extent of pathologist review, ranging from full slide reassessment to review of representative captured images. By performing a retrospective analysis of ∼10,000 HER2 FISH cases scored using breast and gastric criteria, the objective was to define decision boundaries at which technologists' and pathologists' interpretations converge, enabling a data-driven approach to triaging HER2 FISH sign-out. Results showed that agreement between technologists' and pathologists' interpretations was highest when technologists' counts were well beyond the diagnostic cutoffs for the HER2/CEP17 ratio and the number of HER2 signals per cell. Conversely, cases near diagnostic thresholds showed increased discordance, emphasizing the need for full review of the FISH slide in this subset. Individual technologists differed in their percent discordance with pathologists' final results; however, this was not explained by case volume (experience) alone and did not consistently improve over time. In summary, distinct HER2 FISH technologists HER2/CEP17 ratio and HER2 signal number per cell values were identified where concordance with pathologists results exceeded 99%, suggesting cases in these ranges may require only limited pathologist slide review, potentially allowing for more efficient triaging and sign-out of HER2 FISH.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 361 Equivocal Human Epidermal Growth Factor Receptor 2 (HER2) Immunohistochemistry and Fluorescence In Situ Hybridization (FISH) Correlative Study in Invasive Breast Carcinoma: An Institutional Quality Assurance StudyAmerican Journal of Clinical Pathology · 2026
- Inter‐rater agreement of HER2 ‐low scores between expert breast pathologists and the Visiopharm digital image analysis application ( HER2 APP , CE2797 )The Journal of Pathology Clinical Research · 2025
- HER2 testing in multiple solid tumor types: concordance of immunohistochemistry scores between three HER2 scoring algorithmsArchiv für Pathologische Anatomie und Physiologie und für Klinische Medicin · 2026
- Clinicopathological factors associated with HER2 FISH positivity in HER2 IHC 2 + breast cancer: a multicenter studyScientific Reports · 2026
- AI-assisted immunohistochemistry algorithm for enriching HER2-amplified cases among equivocal (2+) breast cancers: a two-cohort proof-of-concept studyBMC Cancer · 2026
- 155 Digital Pathology as an Educational Tool to Improve Concordance in HER2 Equivocal Scoring: Implications for Resident Training and Cost ReductionAmerican Journal of Clinical Pathology · 2026