Patient-ventilator asynchronies beyond the waveform.
Candelaria de Haro, Eduardo Mireles-Cabodevila, Gerard Castells-Tortosa, Leonardo Sarlabous, Lluís Blanch
Critical Care Department, Parc Taulí Hospital Universitari, Institut d'Investigació i Innovació Parc Taulí (I3PT-CERCA), Universitat Autònoma de Barcelona, Sabadell. · Medical Intensive Care Unit, Integrated Hospital Care Institute, Cleveland Clinic, Cleveland, Ohio, USA.
Key points
From the paper's conclusion- Management should move beyond eliminating individual events toward minimizing time outside individualized ranges of acceptable effort and assistance.
Abstract
PURPOSE OF REVIEW: Patient-ventilator asynchrony is commonly treated as a technical mismatch requiring correction, although asynchronous breaths are not uniformly harmful and apparent synchrony may conceal injurious effort. This review summarizes evidence from the past 18 months that redefines asynchrony through inspiratory effort, the transition to assisted breathing, and continuous waveform monitoring. RECENT FINDINGS: Potential harm depends on effort magnitude and timing, breath stacking, ventilatory mode, lung and diaphragm vulnerability, dyspnea, and cumulative exposure. Double triggering, reverse triggering, and ineffective efforts may indicate injury, overassistance, or recovery depending on context. During transition, work shifts from ventilator to patient, exposing patients to both excessive and insufficient effort. Continuous analysis can identify burden, clustering, and physiological trajectories missed by intermittent inspection. Artificial-intelligence models increasingly estimate effort and dynamic driving pressure, but heterogeneous definitions, limited external validation, artifacts, and uncertain actionability constrain implementation. SUMMARY: Management should move beyond eliminating individual events toward minimizing time outside individualized ranges of acceptable effort and assistance. Monitoring should integrate waveforms with effort, lung stress, diaphragm loading, dyspnea, ventilator settings, and temporal context. Artificial intelligence should support bedside reassessment rather than autonomous adjustment. Multicenter validation and physiology-guided trials must establish whether this approach improves patient-centered outcomes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Classification of patient-ventilator asynchronies, development of an updated conceptual framework: Scoping review.Respiratory medicine · 2026
- Patient-ventilator asynchrony during critical care transport: evidence and framework.Frontiers in medicine · 2026
- Physiology-guided mechanical ventilation: Monitoring, proportional assist, and bounded automation.World journal of critical care medicine · 2026
- Artificial Intelligence Algorithm to Monitor Inspiratory Muscle Effort and Patient-Ventilator Dyssynchrony During Mechanical VentilationCritical Care Medicine · 2026
- Challenges and strategies for the identification and correction of patient-ventilator asynchronies in daily practice: a scoping review.Acute and critical care · 2026
- Ventilatory asynchronies induced by routine clinical practices in the intensive care unit: A systematic observation combined with a scoping reviewCanadian Journal of Respiratory Therapy · 2026