Incidence and Clinical Course of Intraoperative Impedance Abnormalities in Cochlear Implantation.
Yoshinori Fujiwara, Nozomu Matsumoto, Noritaka Komune, Teppei Noda, Takashi Nakagawa
Department of Otorhinolaryngology, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan.
Key points
From the paper's conclusion- Most OCs that normalized intraoperatively were first recorded as normal within 11 minutes. OCs persisting beyond this interval may persist postoperatively. Careful monitoring is warranted for SME devices, while decisions regarding device verification or replacement should be made cautiously and require further validation.
Abstract
OBJECTIVE: To investigate the incidence, characteristics, and postoperative course of intraoperative impedance abnormalities during cochlear implant (CI) surgery, with particular attention to features associated with persistent abnormalities. STUDY DESIGN: Retrospective observational study. SETTING: Tertiary referral university hospital. PATIENTS: A total of 324 cochlear implantations were performed between March 2000 and March 2025 with available intraoperative and postoperative impedance measurements. INTERVENTIONS: Cochlear implantation with intraoperative and postoperative electrode impedance measurements. MAIN OUTCOME MEASURES: Incidence of intraoperative impedance abnormalities, time to normalization, postoperative persistence, and device-related differences. RESULTS: Intraoperative impedance abnormalities were detected in 27 of 324 devices (8.3%), involving 102 of 7128 electrode contacts (1.4%). OCs were identified in 25 devices involving 98 electrode contacts, whereas SCs were identified in 2 devices involving 4 electrode contacts. All 12 devices with complete intraoperative resolution had OCs, and all 83 contacts with documented intraoperative normalization were OCs. Among these 83 OC contacts, 80 (96%) were first recorded as normal within 11 minutes. Among the 25 devices with OCs, persistent abnormalities were more frequent in devices with ≤2 affected contacts (66.7%) than in those with ≥3 affected contacts (14.3%) (OR: 12.0; 95% CI: 1.16-123.7; P=0.030). The incidence of abnormalities was higher in the slim modiolar electrode (SME) arrays (8.1%) than in the non-SME arrays (2.8%) (OR: 3.04; 95% CI: 1.05-8.79; P=0.048). CONCLUSION: Most OCs that normalized intraoperatively were first recorded as normal within 11 minutes. OCs persisting beyond this interval may persist postoperatively. Careful monitoring is warranted for SME devices, while decisions regarding device verification or replacement should be made cautiously and require further validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Changes in Electrode Impedance and Cochlear Microphonic in Cochlear Implant Recipients in the First Year After Cochlear Implantation.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Cochlear Implant Electrode Impedances: A 5-Year Study.The journal of international advanced otology · 2026
- Long-Term Impedance Trends Over 10 Years Following Cochlear Implantation: A Retrospective Comparative Analysis.Journal of otolaryngology - head & neck surgery = Le Journal d'oto-rhino-laryngologie et de chirurgie cervico-faciale · 2026
- Association between ECAP signal artifacts and electrical components from a cochlear implant impedance model.Zeitschrift fur medizinische Physik · 2026
- Guiding Cochlear Implant Electrode Insertion Using Impedance Measurements: A Computational Simulation Study.Sensors (Basel, Switzerland) · 2026
- Impedance-based tracking of a cochlear implant electrode reliably detects successful array insertion.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026