Database-optimized versus manufacturer recommended lens constants for the prediction of postoperative refractive error following cataract surgery.
Seth Michel Pantanelli, Luke Rebenitsch, John A Vukich, Sergio Sanchez, Julie M Schallhorn
Department of Ophthalmology, Penn State College of Medicine, Hershey, PA, USA . · ClearSight LASIK and Lens, Oklahoma City, OK, USA.
Key points
From the paper's conclusion- Lens constant optimization using the Veracity surgical planner can result in significant reduction in prediction error. Using these database-optimized lens constants over manufacturer recommended ones may improve refractive outcomes after cataract surgery for most types of IOLs and eyes.
Abstract
PURPOSE: To compare refractive outcomes when Veracity database-optimized (Dopt) or manufacturer-suggested (Mfr) lens constants are used to perform IOL power calculations. DESIGN: Retrospective database study across the United States. SETTING: Database-optimized lens constants (Dopt) were validated in an independent dataset from 268 surgeons. METHODS: This retrospective database validation study analyzed cataract surgery cases planned using Veracity between November 2023 and December 2025. Dopt, previously derived from user-reported outcomes through October 2023, were evaluated against manufacturer-recommended (Mfr) constants in an independent validation dataset. Eligible eyes had their predicted postoperative spherical equivalent (SE) calculated using anterior and posterior keratometry, Barrett Universal II formula, and either Dopt and Mfr lens constants. RESULTS: In total, 14,172 eyes from 14,172 patients were included in this study. Mean prediction error and mean absolute error of the SE was significantly improved with the Dopt constant (-0.004 ± 0.424 D and 0.320 D, respectively) compared to Mfr group (0.167 ± 0.431 D and 0.353 D, respectively, p<0.0001 ). The proportions of eyes with PE within 0.25, 0.50, 0.75 and 1.00 D were also significantly higher for the Dopt constant (50%, 80%, 93% and 97%, respectively) compared to Mfr group (44%, 75%, 91% and 97%, respectively, p<0.0001 ). The use of Dopt constant also resulted in a lower MPE for most subgroups and for 86% of the IOL models assessed in this study. CONCLUSION: Lens constant optimization using the Veracity surgical planner can result in significant reduction in prediction error. Using these database-optimized lens constants over manufacturer recommended ones may improve refractive outcomes after cataract surgery for most types of IOLs and eyes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Segmented Versus Global Optimization of Intraocular Lens Constants Across Axial Length: A Multi-Lens, Multi-Formula Comparison Using a Disjoint Training Dataset.Diagnostics (Basel, Switzerland) · 2026
- Refractive outcomes during transition from ultrasound to optical biometry in cataract surgery: a real-world study from a resource-limited setting.Clinical & experimental optometry · 2026
- Comparison of the Accuracy of the Web-Based ESCRS IOL Power Calculator Formulas.Seminars in ophthalmology · 2026
- Postoperative-Only Estimation of Implanted Intraocular Lens Power Using a Closed-Form Thick-Lens Model With Patient-Specific Predictive Intervals.Translational vision science & technology · 2026
- Formula Discordance and Refractive Target Error: Temporal Validation of Legacy-Modern Intraocular Lens Formula Disagreement as a Quality-Control Marker.Journal of cataract and refractive surgery · 2026
- Stratified Analysis of IOL Constant Optimization and Refractive Prediction Accuracy According to Ocular Biometric Parameters: Comparison with Overall Optimization.Ophthalmology and therapy · 2026