Local Diagnostic Reference Levels for Pediatric Whole-body and Brain [18F]FDG PET/CT.
Théo Assoignon, Michel Hesse, Florian Dupont, Rashel De Stefanis, François Jamar
Institut de Recherche Expérimentale et Clinique, Université Catholique de Louvain, Brussels, Belgium. · Medical Physics Department, Cliniques Universitaires Saint-Luc, Brussels, Belgium.
Key points
From the paper's conclusion- The values reported in this article, by comparison with existing literature, show that it is possible to reduce the administered activity for certain patient categories. Local, national, or international DRLs are expected to change over time and results need to be updated in an iterative way, particularly when new technologies are implemented.
Abstract
BACKGROUND: PET/CT scanners allow anatomical and metabolic imaging but expose patients to ionizing radiation, requiring dose optimization according to the ALARA principle, especially in children. Diagnostic Reference Levels are a possibility, yet no pediatric [18F]FDG PET/CT DRLs exist in Europe. This study reports typical doses for pediatric whole-body and brain [18F]FDG PET/CT. METHODS: A retrospective study included all PET/CT scans performed at our center in patients under 18 y (2014-2018). Patients were grouped by body weight (whole-body: 15≤30 kg, 30≤50 kg, 50≤80 kg) and age (brain: 1≤6 y, 6≤18 y), according to the ICRP 135 classification. DRLs are defined as the 75th percentile (P75) of observed values. RESULTS: For whole-body PET, P75 (median) administered activities were 4.5 (4.4), 4.5 (4.1), and 4.7 (4.3) MBq kg-1 across weight groups. Brain PET P75 (median) values were 5.1 (4.5) MBq kg-1 (1≤6 y) and 4.0 (3.1) MBq kg-1 (6≤18 y). Whole-body CT metrics P75 (median) were: CTDIvol of 2.5 (1.6), 3.1 (2.2), and 3.9 (3.1) mGy; DLP of 211.9 (145.5), 331.0 (230.9), and 464.8 (342.6) mGy·cm. Brain CT showed a P75 CTDIvol of 3.5 mGy for both age groups, with DLPs of 75.8 and 76.2 (median: 75.8) mGy·cm. CONCLUSION: The values reported in this article, by comparison with existing literature, show that it is possible to reduce the administered activity for certain patient categories. Local, national, or international DRLs are expected to change over time and results need to be updated in an iterative way, particularly when new technologies are implemented.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Swiss survey and national diagnostic reference levels for PET/CT examinations in adults.Zeitschrift fur medizinische Physik · 2026
- Optimization of patient dose and image quality for positron emission tomography-computed tomography.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Establishment of preliminary diagnostic reference levels for 18F-FDG PET/CT in adult lymphoma in Costa Rica.Radiation protection dosimetry · 2026
- Local diagnostic reference levels for paediatric computed tomography based on European guidelines: anatomy- and indication-based approaches.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Clinical Determinants of Dose-Length Product in Pediatric Brain CT: Implications for Age-Specific Imaging Protocols.Journal of imaging · 2026
- Estimation of Effective Dose Received by Caregivers During [18F]FDG PET/CT Studies.Journal of nuclear medicine technology · 2026