Development and Validation of an Extended NEMA Body Phantom to Detect 18F-FDG Heterogeneous Distribution in Texture Analysis.
Koichi Okuda, Kotaro Suino, Taisei Komoda, Kazuma Honda, Hiroki Yamamoto, Yuya Watanabe, Shozo Yamashita, Shota Hosokawa,
Department of Radiation Science, Hirosaki University Graduate School of Health Sciences, Hirosaki, Aomori, Japan. · Department of Radiology, Hirosaki University Hospital, Hirosaki, Aomori, Japan.
Key points
From the paper's conclusion- We extended a NEMA phantom by inserting spheres containing beads to represent heterogeneous 18F-FDG distribution.
Abstract
We evaluated heterogeneous 18F-fluorodeoxyglucose (FDG) distribution using texture features and an extended National Electrical Manufacturers Association (NEMA) phantom. Homogeneous and heterogeneous distribution was represented by spheres without or with acrylic beads (5-mm diameter). We reconstructed PET images acquired for 10 min using the Q.clear algorithm and controlled statistical noise by adjusting β values. We calculated maximum (SUVmax) and mean (SUVmean) standardized uptake values and coefficients of variation (COV), then evaluated 22 established texture features to determine homogeneity and heterogeneity. Heterogeneous and homogeneous 18F-FDG distribution was visualized using spherical phantoms with and without beads. The SUVmax and SUVmean were underestimated by ~ 45% in heterogeneous compared with homogeneous 18F-FDG distribution. As the β value increased, the COV for the phantoms with and without beads decreased and eventually became equivalent. When the spherical phantoms with or without beads were imaged using a β value of 100, eight textural features were associated with texture heterogeneity, and six features were associated with texture homogeneity. We extended a NEMA phantom by inserting spheres containing beads to represent heterogeneous 18F-FDG distribution. Robust texture features were determined using the phantom to detect 18F-FDG homogeneity and heterogeneity.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multi-Site Quality Assurance of ADC and PDFF Measurements: A Multi-Vendor Phantom Study.Journal of magnetic resonance imaging : JMRI · 2026
- Impact of noise power spectrum (NPS) peak frequency on CT radiomics features: A computational phantom study.Biomedical physics & engineering express · 2026
- Benchmarking 18F and 68Ga Phantom Performance on a Digital Time-of-Flight PET/CT System: Reference Values for Spatial Resolution, Contrast Recovery, and Background Variability.Journal of nuclear medicine technology · 2026
- Radiomic Profiling of Tumor Heterogeneity Predicts Treatment Response in High-Risk Hodgkin Lymphoma: Results from the Phase III AHOD0831 Trial.Cancers · 2026
- 18F-FDG PET-based radiomics for the prediction of aggressive/high-grade tumor in patients with cartilaginous tumors: A preliminary study.Revista espanola de medicina nuclear e imagen molecular · 2026
- Technical evaluation of delayed-phase SUV estimation using four-phase whole-body dynamic 18F-FDG PET/CT: a phantom and clinical study.Radiological physics and technology · 2026