A Transradial Amputation Procedure to Restore Cutaneous Finger Sensations.
Michael F Fernandez, Mohini L Lutchman, Yibin B Zhang, Christopher C Shallal, Matthew J Carty, Hugh M Herr
From the K. Lisa Yang Center for Bionics, Yang-Tan Collective, McGovern Institute for Brain Research, Massachusetts Institute of Technology, Cambridge, MA. · Department of Neurobiology, Harvard Medical School, Boston, MA.
Key points
From the paper's conclusion- Here, we present an anatomical feasibility and design study for incorporating neurally controlled CMIs into transradial amputation.
Abstract
For the upper extremity, cutaneous afferents play a critical role in grip force modulation and dexterous manipulation. Standard-of-care transradial amputation techniques eliminate these signals and provide no mechanisms for generating physiological tactile hand sensations during manipulation. The cutaneous mechanoneural interface (CMI) is a surgical construct designed to restore such feedback. The CMI consists of a glabrous dermal graft reinnervated by a transected cutaneous or mixed sensory nerve branch. The graft is folded around the transected nerve, and then a muscle actuator circumferentially wraps around the entire nerve-graft construct such that contraction of the actuator strains the dermal graft and activates embedded mechanoreceptors to produce sensation localized to the original receptive field of the donor nerve. Because the graft may contain multiple receptor types and the actuator may admit multiple mechanical stimulus responses, multimodal cutaneous perception is hypothesized for each CMI. In this investigation, we design neurally controlled CMIs, wherein each muscle actuator is natively innervated and vascularized and can be recruited synergistically with the desired CMI cutaneous response. More specifically, we select CMI muscle actuators with volitional activation during grasp such that afferent discharge may be both proportional to, and synchronous with, voluntary effort. This framework is hypothesized to provide neurally modulated, somatotopically matched sensory feedback. Here, we present an anatomical feasibility and design study for incorporating neurally controlled CMIs into transradial amputation. Potential donor sites for skin and nerve, as well as choices of muscle actuators, are discussed to maximize surgical feasibility and functional relevance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The Upper-Limb Residual Limb as a Biological Control Interface: Surgical Preparation, Myoelectric Signal Optimization, Rehabilitation, and Neural Prosthetic Integration—A Narrative ReviewSurgical Techniques Development · 2026
- Spared Motor Neurons Enable Control of a Robotic Sixth Finger for Functional Grasping in Tetraplegia.IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society · 2026
- A wearable reconfigurable surface electrode network for spatiotemporal control of hand motor outputNature Communications · 2026
- Zwischen Mensch und Maschine: Rekonstruktion der oberen ExtremitätWiener klinisches Magazin · 2026
- Mechanonociceptive tactile feedback enhances prosthetic grasp control during repeated human-in-the-loop interaction.Frontiers in neurorobotics · 2026
- Myoelectric Closed Loop Enables Unified Control Across Multiple Modalities of Hand Assistance Using Reconfigurable Robots.IEEE transactions on cybernetics · 2026