P016. Feasibility of brief airway occlusions for measuring upper airway collapsibility

T Vuong, K Reynolds, T Pang, A Vakulin, T Farook, T Nguyen, D Nguyen, P Catcheside

Key points

From the paper's conclusion
  • Discussion More data are required, but limited preliminary results support the potential for brief upper airway occlusions to help simplify awake assessments of upper airway collapsibility.

Abstract

Introduction Obstructive sleep apnoea (OSA) negatively impacts sleep, health and quality of life. Although OSA has different causal endotypes, most patients exhibit abnormal upper airway collapsibility, which is useful to measure to guide treatment choices. One approach is to measure the upper airway collapsibility index (UACI), via negative pressure pulses (NPP) awake. However, NPP delivery is technically complex. This study proposes simple brief airway occlusions as an alternative to NPP. Methods In this ongoing study, ten participants are fitted with a nasal mask to measure airflow and mask pressure, along with airway catheters to record choanal and epiglottic pressures. Brief replicate (~20) NPP or occlusions (250 ms) are delivered to the airway during early-mid inspiration via a flow-triggered solenoid system during wake. Results Initial findings from the first study participant support the feasibility of brief airway occlusions from which to evaluate UACI and other metrics of airway collapsibility. Peak negative choanal pressures delivered with occlusions appeared to be more consistent versus NPPs. Discussion More data are required, but limited preliminary results support the potential for brief upper airway occlusions to help simplify awake assessments of upper airway collapsibility.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers