Are different consumer sleep technologies measuring the same essential aspects of sleep?
Kiran K G Ravindran, Ciro Della Monica, Giuseppe Atzori, Marta M Pineda, Ramin Nilforooshan, Hana Hassanin, Victoria Revell, Derk-Jan Dijk
UK Dementia Research Institute, Care Research and Technology Centre at Imperial College, London, and University of Surrey, Guildford, UK.
Key points
From the paper's conclusion- Correlation of most sleep measures across assessment modalities is poor, suggesting that the consumer sleep technologies tested in our study may not measure the same essential aspects of sleep. Duration was the only sleep aspect transferable between these devices.
Abstract
STUDY OBJECTIVES: Consumer sleep technologies (CSTs) enable low-burden longitudinal sleep monitoring, and their measures are often interpreted as equivalent to polysomnography (PSG) measures. We applied measurement reliability-aware approaches to determine whether CST-derived 'sleep' measures (1) are interchangeable or device-specific, (2) can reliably assess participant-level sleep characteristics and (3) converge onto interpretable principal components representing essential aspects of sleep. METHODS: Data from 74 older adults (20 people living with dementia) were collected at-home (total=752 paired nights) using four modalities simultaneously: research-grade actigraphy (Axivity), a wearable (Withings Watch), a nearable (Withings Sleep Analyser) and Sleep Diary, followed by one in-lab PSG assessment. We applied Bland-Altman analyses, repeated-measures correlation analyses, intraclass correlation coefficients (ICC) and principal component analysis (PCA). RESULTS: Single-night between-device correlations and correlations with PSG were weak (r<0.3) for most measures but moderate (0.3≤r<0.7) for duration- and timing measures. Bland-Altman analyses revealed systematic disagreement with PSG, with total sleep time and sleep efficiency overestimated, wake after sleep onset underestimated, and poor agreement for sleep-stage durations. The number of nights required to obtain acceptable reliability (ICC≥0.7) varied across measures, modalities and participant groups, with 71% of measures reaching an ICC ≥0.7 within seven nights. Reliability-filtered PCA yielded interpretable principal components, with Duration being the only sleep aspect showing consistent moderate between-device associations. CONCLUSIONS: Correlation of most sleep measures across assessment modalities is poor, suggesting that the consumer sleep technologies tested in our study may not measure the same essential aspects of sleep. Duration was the only sleep aspect transferable between these devices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- P013. Concordance in sleep measurement between a home test and PSGSLEEP advances · 2026
- Comparing wearable and nearable devices for sleep monitoring: A within-subject comparison of wrist-worn wearable and radar-based nearable devices for naturalistic home-based sleep monitoring.Digital health · 2026
- Performance evaluation of consumer sleep-tracking wearables and nearables in healthy young and older adultsSLEEP Advances · 2026
- KI und Wearables: Neue Wege für Schlaf, Gesundheit und Wellbeingpsychopraxis. neuropraxis · 2026
- Comparing the reliability of sleep metrics across two wrist actigraphs: Philips Actiwatch Spectrum Plus and ActiGraph CentrePoint Insight watchSleep Science and Practice · 2026
- Sensors and computational methods for sleep quality assessment: A systematic reviewResults in Engineering · 2026