Real-world performance of a Food and Drug Administration-approved artificial intelligence tool for mammography in patients with prior breast surgery or indwelling biopsy clips.
Emily B Ambinder, Colin Paulbeck, Shuchi Zinzuwadia, Lisa Mullen, Kelly S Myers, Eniola Oluyemi, Babita Panigrahi
Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, MD 21287, United States.
Key points
From the paper's conclusion- Purpose To evaluate the real-world performance of an FDA-approved AI mammography tool in patients with prior breast surgery or biopsy clips.
Abstract
BACKGROUND: Artificial intelligence (AI) tools are increasingly used in clinical screening mammography, but their performance in patients with prior breast surgery or indwelling biopsy clips, who represent a substantial proportion of screening populations, has not been well characterized. PURPOSE: To evaluate the real-world performance of an Food and Drug Administration-approved AI mammography tool in patients with prior breast surgery or biopsy clips. METHODS AND MATERIALS: We retrospectively reviewed all screening mammograms performed at our institution between August 1, 2024, and February 28, 2025, after implementation of a commercially available AI tool (Transpara 1.7.4-A), merging clinical outcomes with the AI risk score (low, intermediate, or elevated). Patients were stratified by procedural history into 3 groups: prior breast surgery, biopsy clip, and reference (ie, neither intervention). Sensitivity, positive predictive value (PPV), and false positive rate (FPR) at the elevated and intermediate + elevated thresholds were compared using Fisher exact tests presented as percentage (95% confidence interval [CI]). RESULTS: A total of 24 427 screening mammograms were included (mean age 61 ± 13 years): 19 287 in the reference group, 2647 in the surgery group, and 2493 in the biopsy clip group. AI sensitivity at the intermediate + elevated threshold was preserved (reference 92.1% [95% CI, 86.9%-95.3%], surgery 90.5% [95% CI, 71.1%-97.3%], biopsy clip 90.0% [95% CI, 69.9%-97.2%]; both P > .6). The AI-PPV of an elevated exam was correspondingly reduced in altered-breast patients (reference 13.6% [95% CI, 11.1%-16.6%], surgery 5.5% [95% CI, 2.9%-10.0%], biopsy clip 4.9% [95% CI, 2.5%-9.4%]; both P < .005 vs reference). The AI-FPR was higher in both subgroups at the intermediate + elevated threshold (reference 35.6% [95% CI, 34.9%-36.3%] vs surgery 47.9% [95% CI, 46.0%-49.9%] and biopsy clip 47.9% [95% CI, 45.9%-49.8%]; both P < .001) and more than doubled at the elevated threshold (reference 2.7% [95% CI, 2.5%-2.9%] vs surgery 5.9% [95% CI, 5.1%-6.9%] and biopsy clip 6.2% [95% CI, 5.3%-7.2%]; both P < .001). CONCLUSION: AI examination-level sensitivity for breast cancer was preserved in patients with prior breast surgery or indwelling biopsy clips. However, an elevated AI risk score was less likely to reflect a cancer in these patients than in the reference group, with a lower AI-PPV and higher AI-FPR.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance of an Artificial Intelligence Support System on Screening Mammography Cases Proceeding to Stereotactic BiopsyCancers · 2025
- AI-based triage and decision support in mammography and digital tomosynthesis for breast cancer screening: a paired, noninferiority trialNature Medicine · 2026
- Comparison of Artificial Intelligence Detection and Short-Term Breast Cancer Risk Models within and beyond Their Intended Use in Mammography Screening.Radiology. Imaging cancer · 2026
- Diagnostic performance of FDA-cleared and CE-marked AI systems for mammography screening: a systematic review and meta-analysis.European journal of radiology · 2026
- Diagnostic accuracy, fairness and clinical implementation of AI for breast cancer screening: results of multicenter retrospective and prospective technical feasibility studiesNature Cancer · 2026
- Artificial Intelligence as triage in breast screening: simulated impact on workload and detectionEuropean Journal of Public Health · 2026