Empowering Movement: A Human-in-the-Loop Proportional Controller for Assisted Elbow Flexion in Brachial Plexus Injury.
Sandesh G Bhat, Alexander Y Shin, Kenton R Kaufman
Motion Analysis Laboratory, Department of Orthopedic Surgery, Mayo Clinic, 200 First St. SW, Rochester, MN 55902, United States. · Division of Hand Surgery, Department of Orthopedic Surgery, Mayo Clinic, 200 First St. SW, Rochester, MN 55902, United States.
Key points
From the paper's conclusion- A low-powered office computer was sufficient to train an NNR with training time short enough to be completed in an orthotist's office within a typical appointment time. These data verify that the process can be field deployable.
Abstract
INTRODUCTION: Although powered orthotic devices are becoming more available and increasingly recognized by the medical community, several factors still prevent them from being widely adopted. These orthoses typically use a simple controller, often termed a "bang-bang" controller, which actuates a motor if the muscle's electromyography (EMG) signal is above a set threshold. A proportional controller that could actuate the elbow mechanism based on the EMG signal magnitude would enable a better human-machine connection. The real challenge for designing a proportional controller lies in the heterogeneity of the types of injury and surgeries used in this patient population, as well as the extent of recovery post-surgery. Neural networks effectively model nonlinear relationships in control system design. Recently, a novel powered myoelectric elbow orthosis (PMEO) was developed and tested on patients with a brachial plexus injury (BPI) that used a bang-bang controller. The report evaluated the possibility of using a proportional controller using a neural network-powered regression (NNR) for this exoskeleton. MATERIALS AND METHODS: Data from 31 participants with a BPI were collected after receiving IRB approval. A custom apparatus was designed to measure elbow flexion torque and EMG, while participants were instructed to match their torque to a predefined target (10%-40% of MVC). A 2-layered NNR (2 hidden nodes) with a Tanh activation and 5 cross-validation folds was used to find a relationship between the properties of the EMG (moving average, root mean square, variance, standard deviation, and slope) and the elbow torque for individual participants. The collected data were divided into a training set (70%) and a test set (30%). The root mean squared error (RMSE) for the test set was calculated for each NNR model, along with the training time and prediction speed. RESULTS: A standard computer (Intel Core i5-10500 with 16 GB RAM) was used to train the models. The trained models required 82 seconds (range: 3-332 seconds) to train on 8.2e4 (range: 7.7e4-8.9e4) samples. The test dataset consisted of 3.5e4 (range: 3.3e4-3.8e4) samples. The median test RMSE was 0.39 Nm (range: 0.1-3.09 Nm). The NNR took 7 milliseconds (range: 6-10 milliseconds) to predict the test dataset; 35% and 16% of the models had an R2 > 0.5 and >0.7, respectively. CONCLUSIONS: A low-powered office computer was sufficient to train an NNR with training time short enough to be completed in an orthotist's office within a typical appointment time. These data verify that the process can be field deployable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Neuro-adaptive predictive control for upper-limb rehabilitation using EMG and intention-aware assistance.Medical & biological engineering & computing · 2026
- Comparing volitional electromyographic control strategies: from proportional mapping to muscle modeling and reflex augmentation.Biomedical engineering letters · 2026
- A hybrid EMG–EEG interface for robust intention detection and fatigue-adaptive control of an elbow rehabilitation robotScientific Reports · 2025
- Robotic Training Using a Novel Upper-Limb Hybrid Assistive Limb After Brachial Plexus Injury: An Electrophysiological Observational Case Series Study.Journal of functional morphology and kinesiology · 2026
- Intelligent observer-enhanced adaptive robust controller of a lower-limb rehabilitation exoskeleton driven by pneumatic artificial muscles.ISA transactions · 2026
- User Intention Estimation for an Elbow Exoskeleton using Surface Myoelectric Pattern RecognitionCurrent Directions in Biomedical Engineering · 2026