Assessment of a chemiluminescence immunoassay for antimitochondrial antibody subtype M2 in the diagnosis of primary biliary cholangitis.
Yujiao Jin, Ying Wang, Aifang Xu, Miaochan Wang, Kenv Pan
Department of Clinical Laboratory, Hangzhou Xixi Hospital, Hangzhou, China.
Key points
From the paper's conclusion- AMA-M2 detected by chemiluminescence immunoassay had a sensitivity of 84.1%, a specificity of 90.1%, and an accuracy of 88.4%. Our receiver operating characteristic curve analysis showed an area under the curve of 0.927 (95% CI, 0.903-0.951) for AMA-M2 by chemiluminescence immunoassay, which was higher than the area under the curves for AMA-M2 by LIA (0.772) and M2-3E by LIA (0.862) (both P < .001). High concordance was observed between AMA-M2 by chemiluminescence immunoassay and M2-3E by LIA (96.1%, κ = 0.908), with a strong correlation between the 2 assays.
Abstract
INTRODUCTION: We sought to evaluate the diagnostic performance of and interassay agreement between a chemiluminescence immunoassay for antimitochondrial antibody subtype M2 (AMA-M2) in patients with primary biliary cholangitis and clinically relevant disease controls with other liver diseases. METHODS: In this retrospective study, 164 patients with primary biliary cholangitis and 403 patients with other liver diseases were enrolled between December 2020 and January 2023. The AMAs were detected using indirect immunofluorescence, line immunoassay (LIA), and chemiluminescence immunoassay. Diagnostic performance was evaluated by sensitivity, specificity, accuracy, and receiver operating characteristic curve analysis. Agreement between methods was assessed using the Cohen κ coefficient and correlation analysis. RESULTS: AMA-M2 detected by chemiluminescence immunoassay had a sensitivity of 84.1%, a specificity of 90.1%, and an accuracy of 88.4%. Our receiver operating characteristic curve analysis showed an area under the curve of 0.927 (95% CI, 0.903-0.951) for AMA-M2 by chemiluminescence immunoassay, which was higher than the area under the curves for AMA-M2 by LIA (0.772) and M2-3E by LIA (0.862) (both P < .001). High concordance was observed between AMA-M2 by chemiluminescence immunoassay and M2-3E by LIA (96.1%, κ = 0.908), with a strong correlation between the 2 assays. DISCUSSION: Chemiluminescence immunoassay-based detection of AMA-M2 showed high diagnostic performance and strong agreement with conventional assays in this cohort, supporting its utility as a reliable quantitative method for clinical diagnosis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Letter: Double PBC ‐Related Autoantibody Positivity—How Essential Is It for Diagnosis?Alimentary Pharmacology & Therapeutics · 2026
- Clinical application of serum Mac-2-binding protein glycosylation isomer in the diagnosis of liver fibrosis: a meta-analysis.Cancer treatment and research communications · 2026
- Dual Positivity for AMA / AMA ‐ M2 and Anti‐gp210/sp100 Shows Highest Diagnostic Value for Primary Biliary CholangitisAlimentary Pharmacology & Therapeutics · 2026
- Case Report: persistent diaphragmatic and cardiac involvement despite creatine kinase normalization in anti-mitochondrial M2 antibody-positive inflammatory myopathy with primary biliary cholangitis.Frontiers in immunology · 2026
- Isotope-Coded Click Chemistry-Enabled Enrichment for Reliable Profiling and Quantification of Carboxyl Metabolites.Analytical chemistry · 2026
- Clinical utility of serum Mac-2 binding protein glycosylation isomer as a noninvasive marker for liver fibrosis in biliary atresia.Journal of pediatric surgery · 2026