Validation of a point-of-care assay for simultaneous quantification of equine adrenocorticotropic hormone and corticotropin-like intermediate peptide.

Kelsey A Hart, Londa J Berghaus, Ian Harmon

Department of Large Animal Medicine, University of Georgia College of Veterinary Medicine, Athens, GA 30602, United States. · Zomedica®, Inc; Plymouth, MN, United States.

Key points

From the paper's conclusion
  • Conclusions and clinical importance These data demonstrate acceptable field performance for the POC-ACTH/CLIP assay and good agreement with the chemiluminescent iACTH immunoassay, with a larger bias in samples with higher hormone concentrations.

Abstract

BACKGROUND: Plasma adrenocorticotropic hormone (ACTH) concentrations are used to diagnose equine pituitary pars intermedia dysfunction (PPID). The reference chemiluminescent immunoassay for immunoreactive ACTH (iACTH) cross-reacts with corticotropin-like intermediate peptide (CLIP), which is also abnormally high in PPID. HYPOTHESIS/OBJECTIVES: To compare the performance of a novel point-of-care (POC) immunoassay that concurrently measures ACTH and CLIP against the reference iACTH assay. ANIMALS: Twenty-eight university-owned horses (13 PPID, 15 non-PPID). METHODS: Prospective observational study. Resting plasma hormone concentrations in all animals were measured by both methods in all seasons over 1 year, and after thyrotropin releasing hormone (TRH)-stimulation testing once. Bias of the POC-ACTH/CLIP assay was assessed using Spearman correlation analysis and Bland-Altman plots, and precision and linearity of the POC-ACTH/CLIP assay were determined. RESULTS: For resting and TRH-stimulated ACTH concentrations, the POC-ACTH/CLIP assay and the reference method had good agreement overall (r ≥ 0.94; P < .001) and in PPID animals (r = 0.93, P < .001). Bias was greater in autumn (33.6-35.6 ± 89.5-96 pg/mL) compared to winter/spring (1.6-11.1 ± 8.4-29.6 pg/mL) and in samples with ACTH concentrations ≥ 90 pg/mL (87.8 ± 116.4 pg/mL). Intra- and inter-assay coefficients of variation for the POC-ACTH/CLIP assay were 8.47% and 14.1%, respectively, and assay linearity was excellent (R2 = 0.991). CONCLUSIONS AND CLINICAL IMPORTANCE: These data demonstrate acceptable field performance for the POC-ACTH/CLIP assay and good agreement with the chemiluminescent iACTH immunoassay, with a larger bias in samples with higher hormone concentrations.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Validation of a point-of-care assay for simultaneous quantification of equine adrenocorticotropic hormone and corticotropin-like intermediate peptide — Science Skim