Validation of a point-of-care assay for simultaneous quantification of equine adrenocorticotropic hormone and corticotropin-like intermediate peptide.
Kelsey A Hart, Londa J Berghaus, Ian Harmon
Department of Large Animal Medicine, University of Georgia College of Veterinary Medicine, Athens, GA 30602, United States. · Zomedica®, Inc; Plymouth, MN, United States.
Key points
From the paper's conclusion- Conclusions and clinical importance These data demonstrate acceptable field performance for the POC-ACTH/CLIP assay and good agreement with the chemiluminescent iACTH immunoassay, with a larger bias in samples with higher hormone concentrations.
Abstract
BACKGROUND: Plasma adrenocorticotropic hormone (ACTH) concentrations are used to diagnose equine pituitary pars intermedia dysfunction (PPID). The reference chemiluminescent immunoassay for immunoreactive ACTH (iACTH) cross-reacts with corticotropin-like intermediate peptide (CLIP), which is also abnormally high in PPID. HYPOTHESIS/OBJECTIVES: To compare the performance of a novel point-of-care (POC) immunoassay that concurrently measures ACTH and CLIP against the reference iACTH assay. ANIMALS: Twenty-eight university-owned horses (13 PPID, 15 non-PPID). METHODS: Prospective observational study. Resting plasma hormone concentrations in all animals were measured by both methods in all seasons over 1 year, and after thyrotropin releasing hormone (TRH)-stimulation testing once. Bias of the POC-ACTH/CLIP assay was assessed using Spearman correlation analysis and Bland-Altman plots, and precision and linearity of the POC-ACTH/CLIP assay were determined. RESULTS: For resting and TRH-stimulated ACTH concentrations, the POC-ACTH/CLIP assay and the reference method had good agreement overall (r ≥ 0.94; P < .001) and in PPID animals (r = 0.93, P < .001). Bias was greater in autumn (33.6-35.6 ± 89.5-96 pg/mL) compared to winter/spring (1.6-11.1 ± 8.4-29.6 pg/mL) and in samples with ACTH concentrations ≥ 90 pg/mL (87.8 ± 116.4 pg/mL). Intra- and inter-assay coefficients of variation for the POC-ACTH/CLIP assay were 8.47% and 14.1%, respectively, and assay linearity was excellent (R2 = 0.991). CONCLUSIONS AND CLINICAL IMPORTANCE: These data demonstrate acceptable field performance for the POC-ACTH/CLIP assay and good agreement with the chemiluminescent iACTH immunoassay, with a larger bias in samples with higher hormone concentrations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Plasma adrenocorticotropin concentrations vary with assay used in horses with and without pituitary pars intermedia dysfunction.Domestic animal endocrinology · 2026
- Evaluation of a new immunoassay for measuring equine serum insulin and plasma adrenocorticotrophic hormone: A method comparison study.Equine veterinary journal · 2026
- Preliminary evaluation of beta-endorphin alone and in combination with adrenocorticotropic hormone for the diagnosis of pituitary pars intermedia dysfunction in equids.Journal of veterinary internal medicine · 2026
- Comparison of plasma separation techniques prior to freezing on equine immunoreactive adrenocorticotrophic hormoneEquine Veterinary Education · 2026
- Pituitary Pars Intermedia Dysfunction.The Veterinary clinics of North America. Equine practice · 2026
- Sedated dental procedures increase endogenous plasma ACTH concentration in horses in the fall.Journal of the American Veterinary Medical Association · 2026