Journal of burn care & research : official publication of the American Burn Association

Skin Graft Thickness Consistency: A Comparison of Traditional Dermatome and Rotating Circular Blade Technology.

Timothy Worley, Alexandra McDowell, Michelle Tarbox, Elizabeth Brown, John Griswold, Alan Pang, Debra Kurtz

Texas Tech University Health Sciences Center, School of Medicine, Lubbock, TX, USA. · University Medical Center, Lubbock, TX, USA.

Key points

From the paper's conclusion
  • These findings suggest that device design may mitigate operator-dependent variability and improve reproducibility in split-thickness skin graft harvesting.

Abstract

Our objective was to evaluate the accuracy and intra-graft consistency of split-thickness skin graft thickness using a traditional dermatome versus a dermatome with a rotating circular blade. A controlled cadaveric study was conducted with operators across four experience levels. Each participant harvested split-thickness skin grafts using both devices set to 0.009 inches (0.229 mm). Grafts were obtained from adjacent locations on the back and leg. Three 6-mm punch biopsies per graft were analyzed histologically for thickness. Outcomes included deviation from the depth setting and intra-graft variability. Statistical comparisons included independent t-tests and analysis of variance. A total of 118 samples were analyzed. Traditional dermatome grafts were significantly thicker than intended (mean 0.0119 inches (0.30 mm); +33%), whereas grafts taken with the rotating circular blade dermatome were closer to target (mean 0.0080 inches [0.20 mm]; -11%) (p = 0.046). Intra-graft variability was significantly lower with the rotating circular blade dermatome (F(1,118) = 26.75, p < 0.001). No significant differences were observed by operator experience or donor site. The rotating circular blade dermatome demonstrated superior depth accuracy and reduced intra-graft variability compared with traditional dermatome. These findings suggest that device design may mitigate operator-dependent variability and improve reproducibility in split-thickness skin graft harvesting.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Skin Graft Thickness Consistency: A Comparison of Traditional Dermatome and Rotating Circular Blade Technology — Science Skim